1 Calibration Instruction for Salwico ST400 H2S
Calibration Instruction
2
5103028_Salwico ST400 H2S_Calibration Instruction_I0_EN_2019_C
A
Test points
B
SPAN potentiometer
C
ZERO potentiometer
1.2 Prerequisites
1
Use the arrow keys on the control panel to
manoeuvre through the different menus.
Press the
right arrow key to access the
menu tree.
2
Scroll to
4 Gas Detectors
and press
Continue
.
3
Scroll to the detector that should be
calibrated and press
Options
.
4
Scroll to
3 Disable
and choose
Continue
.
5
You have to confirm the disablement before
you can proceed.
Press
Disable
to continue.