Concoa 308 Series Installation And Operation Instructions Manual Download Page 8

8

Trouble Shooting

Symptoms

1. With no flow through the 

system (downstream valves 

closed and cylinder valve 

open) working pressure 

increasing steadily above set 

pressure (use  test gauge if 

necessary).

2.  Gas leakage from spring case

Gauge:

3.  Gas leakage from gauge

4.  Gauge does not return to zero

5.  Gauge does not consistently 

repeat reading

6.  Gas leakage from any joint

7. Excessive drop in working 

pressure

8.  Gas leakage from relief valve

9.  Inadequate heat

10. No heat

Probable Cause

1.  Leak across the seat

2.  Loose case or damaged 

diaphragm

3.  Damaged gauge

4.  Damaged gauge

5.  Damaged gauge

 

6.  Damaged or loose connection 

(fittings, gauges, or spring case)

7.  Worn or sticking internal parts;

  Internal flow obstruction; 

Dirty inlet gland filter; Cylinder 

or upstream valve not fully 

opened.

8.   Faulty valve; Seat leak

9.   One or two heaters burned out*

10.  Lack of power; Faulty 

thermostat(s)

*Check resistance across power leads (Figure1).  With all heaters 

functioning, resistance should be 88 ohms.  With two heaters functioning, 

resistance should be 132 ohms.  And with only one heater functioning, 

resistance should be 256 ohms.

Summary of Contents for 308 Series

Page 1: ...l Certified ISO 9001 ADI 3005 D Controls Corporation of America 1501 Harpers Road Virginia Beach VA 23454 To Order Call 1 800 225 0473 or 757 422 8330 Fax 757 422 3125 www concoa com Before Installing...

Page 2: ...2 This Page Intentionally Left Blank...

Page 3: ...maintenance damage improper repair or alteration by anyone other than CONCOA or a service facility designated by CONCOA CUSTOMER SERVICE In the event of equipment failure call CONCOA Customer Service...

Page 4: ...e regulators These regulators have metal diaphragms which make them ideal for general laboratory uses as well as pressure control devices for critical analytical procedures Generally these regulators...

Page 5: ...fully and the required gas flow is ready for use f Close the cylinder valve and unplug the line cord from the supply receptacle whenever the regulator is not in use g Before making any electrical conn...

Page 6: ...ont In case of malfunction returnRegulatorforrepairtothenearestCONCOAdesignatedrepair location as special tools tests and techniques are required CAUTION Itisrecommendedthatallservicingbedonebyaservic...

Page 7: ...estic BLACK WIRING DIAGRAM or BROWN or GREEN YELLOW BLUE or Secondary Primary Relief Valve INLET OUTLET Thermostat Thermostat WHITE WHITE WHITE BLACK GREEN or GREEN YELLOW BLACK 115 V Domestic 115 V D...

Page 8: ...quate heat 10 No heat Probable Cause 1 Leak across the seat 2 Loose case or damaged diaphragm 3 Damaged gauge 4 Damaged gauge 5 Damaged gauge 6 Damaged or loose connection fittings gauges orspringcase...

Page 9: ...9 This Page Intentionally Left Blank...

Page 10: ...10 This Page Intentionally Left Blank...

Page 11: ...han Controls Corporation of America or if the equipment has been subject to abuse misuse negligence or accident Controls Corporation of America s sole and exclusive obligation and Buyer s sole and exc...

Page 12: ...Certified ISO 9001 ADI 3005 D Controls Corporation of America 1501 Harpers Road Virginia Beach VA 23454 To Order Call 1 800 225 0473 or 757 422 8330 Fax 757 422 3125 www concoa com...

Reviews: