18
Table 2: Electromagnetic immunity 1
Guidance and manufacturer's statement - Electromagnetic immunity
Comper SkinEnergy is intended for use in the electromagnetic environment as
specified below. The buyer or user should assure that it is used in such an
environment:
Immunity test
IEC 60601 Test Level
Compliance
level
Electromagnetic
environment -
Guidance
Electrostatic
discharge
IEC61000-4-2
±6 kV, contact discharge
±8 kV, air discharge
±6 kV, contact
discharge
±8
kV,
air
discharge
Floors
shall
be
wood, concrete or
ceramic
tile.
If
floors are covered
with the synthetic
material,
the
relative humidity
should be at least
30%.
Electrical
fast
transient burst
IEC61000-4-4
±2 kV, to power supply
lines
±1 kV, to input/output
lines
Not applicable
Not applicable
Surge
IEC61000-4-5
± 1 kV differential mode
voltage
± 2 kV, common mode
voltage
Not applicable
Not applicable
Voltage
dip,
short
interruption and
voltage variation
on power supply
input lines
IEC61000-4-11
<5% UT, for 0.5 cycles
(>95% dip in UT)
40% UT, for 5 cycles
(60% dip in UT)
70% UT, for 25 cycles
(30% dip in UT)
<
5% UT, for 5s
(>95% dip in UT)
Not applicable
Not applicable
Power frequency
magnetic
field
3A/m
3A/m,50/60Hz
The
power-
frequency
13