17
of Comper SkinEnergy. Therefore during use, the product should be kept away
from electromagnetic interference resources, such as radio station and microwave
ovens, etc.
;
See details of guidance and manufacturer's statement in the subsequent
“Guidance and Manufacturer's Statement”
.
Warnings:
Comper SkinEnergy shall not be used adjacent to or stacked with other equipment,
and if adjacent or stacked use is necessary, it shall be verified the device can work
in normal operation in the configuration in which it will be used;
If the physiological parameters measured by the equipment are lower than the
specified minimum values, the measurement results may be inaccurate.
Table 1: Electromagnetic emission
Guidance and manufacturer's statement - Electromagnetic emission
Comper SkinEnergy is intended for use in the electromagnetic environment as
specified below. The buyer or user of Comper SkinEnergy should assure that it is
used in such an environment:
Emission test
Compliance
Electromagnetic environment - Guidance
RF emission
CISPR 11
Group 1
Comper SkinEnergy uses RF energy only for its
internal function. Therefore, RF emissions are
very low and are not likely to cause any
interference to nearby electronic equipment.
RF emission
CISPR 11
Class B
Comper SkinEnergy is suitable for use in all
establishments,
including
domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic emission
IEC61000-3-2
Not
applicable
Voltage
fluctuation/flicker
emission
IEC61000-3-3
Not
applicable