Chiller Unit Storage:
1.
Turn off chiller unit, disconnect the power cord.
2.
Disconnect the extension tubing.
3.
Empty the circulation water from chiller unit by open bottom drain cap, then clean and
disinfect the reservoir, then tight the cap after open dry.
4.
Pack the unit for storage.
Restart: after long term storage, please fill in clean water with less degerming agent, run system
over 10 minutes with cooling vest connection. Then empty the system and re-fill in the liquid as
need.
Trouble shoot and Fault Code
Item
Code
Description
1
E1
Wrong Polarity or reverse connection from power input
2
E2
Low voltage protection, less than 18V
3
E3
High voltage protection, more than 32V
4
P1
No liquid temp signal from sensor
Cautions:
1.
Please confirm chiller voltage is 110V or 220V before connect power.
2.
Please do not turn ON/OFF to run to stops system frequently within a short period, it may
affect the refrigeration system, and significant enhance power consumption.
3.
Please do not block air inlet and outlet, it may cause the problem of compressor overheat or
less cooling capacity.
4.
Please make sure cooling garments or vest be connected before start pump circulation, it may
cause unit leak because of pump pressure.
5.
Please use anti-freeze liquid if temperature setting lower than 0
℃
6.
Please do not run the system close to fire or under water.
7.
Max operation temp is 45
℃
, unit could be overheated protection in hot ambient.
8.
Please stop operation and disconnect the power if high vibration or abnormal noise.