4.
Temp setting: user may press up and down to set up the temp for circulation liquid as need.
5.
Start circulation:
User may press the pump to start or stop circulation. Please make sure the tubing is connected
(no kinks) before circulation.
Clean and Maintenance
Chiller Unit:
Unplugged the power, use a clean damp cloth to clean the outside of chiller unit.
Use compressed air to remove dust and debris.
Reservoir cleaning
Please disconnect the extension tubing from chiller unit.
Empty the remain water or liquid from drain cap on the bottom, and open for dry.
Condenser clean
To keep the system at optimum cooling capacity, the condenser should be kept free of dust and
dirt, user may check and clean it if necessary.
Open the side panel and remove the fans, use 50-100psi compressed air to clean the
contamination.
Charging Refrigerant
(not recommend for uncertified operator)
If cooling capacity was decreased because of lack of refrigerant, user may vacuum the unit and
recharge 2000g R134a refrigerant.
Liquid Heating & Cooling Pad or garment:
Machine wash, wash liquid heating and cooling pad using a front-loading wash machine with cold
water on a gentle/delicate cycle, secure the connection tubes to minimize risk of it flailing in the
spin cycle and damaging connector and tubing sewing.
Note: DO NOT BLEACH, NO IRON, NO DRYER, TUMBLE DRY ON LOW