Satin-Air APP — Operation Manual
Pressure Adjustment
It is recommended that the pressure-adjustment knob is turned to firm when
the bubble pad is inflated for the first time. Users can then easily adjust the
pressure level of the bubble pad to a desired firmness or according to the
suggestion from a health care professional.
If the pressure is consistently low, check for any leakage occurring in the in-
flation tubes/pad. Contact your local dealer for replacement of any damaged
tubes/pad.
For Patients:
WARNING
•
DO NOT
use another pump other than the Satin-Air APP Pump supplied
with this bubble pad. This may result in bubble-puncture.
•
DO NOT
change any component by yourself. If replacement or repair
is needed, contact your local dealer or service center.
CLEANING
This section describes the procedures to clean and decontaminate the
APP system. It is important to follow these procedures before using the
system on patients.
Pump
•
BEFORE
cleaning, unplug the system from the electrical socket.
•
DO NOT
immerse or soak pump in water.
• Check for external damage and move the pump to clean, flat and level
work surface where you will be able to clean your Satin-Air APP Pump.
• Place the pump on the work surface
• Spray a cloth with a quaternary ammonium solution first and wipe the
outside of the case and faceplate.
•
DO NOT
spray any cleaning solution directly on the surface.
•
DO NOT
use a Hypocarbonate or Phenolic based cleaning solution
as this may cause damage to the case.
•
DO NOT
allow excess cleaning solution on faceplate or control panel.
(If solution gets inside the pump, damage will occur.)
• Wipe with a clean cloth and make sure all areas are clean and dry
(top, bottom and both sides).
• After the pump is thoroughly cleaned and dried, proceed to plug in the
pump and test to see if it runs normally.
• Unplug the pump and store with proper identification tag.