![CommScope FIBER INDEXING TERMINAL Installation Instructions Manual Download Page 27](http://html.mh-extra.com/html/commscope/fiber-indexing-terminal/fiber-indexing-terminal_installation-instructions-manual_2643637027.webp)
Page 27 of 29
© 2017 CommScope, Inc. All Rights Reserved
14.2. Cleaning DLX connector and adapter
14.2.1. DLX connector cleaning
14.2.1.1. Clean both the optical port adapter and the
drop cable connector (requires accessory kit FHD-ACC-
CLNKIT1) as specified in ADCP-96-163.
14.2.1.2. Examine the end of the drop cable connector,
note the position of the protrusions on the connector.
14.2.1.3. Locate the cleaning tape cassette that is
provided with the cleaning kit.
14.2.1.4. Open the tape shutter by squeezing the lever
on the underside of the cassette and then keep the shut-
ter open by continuing to squeeze the lever.
14.2.1.5. Hold connector facing you with the protrusions
perpendicular to the length of the cleaning cassette, tilt
the connector to the right (no more than 8-degrees) this
places the connector in the correct position for cleaning.
Note:
The drop cable connector uses angled SC type
connectors. Make sure the connector end-face is point-
ing in the correct direction before starting the cleaning
motion
.
14.2.1.6. With light pressure, slide the connector end-
face once across the tape in the direction shown using a
smooth linear motion. Do not press too hard and do not
repeat the cleaning motion with the same tape.
14.2.1.7. Release the lever on the underside of the cas-
sette to close the tape shutter.
14.2.1.8. Repeat Step 11.4.2.3 through Step 11.4.2.6 until
the connector has been cleaned three times.
14.2.1.9. When the connector and the adapter are clean,
install the connector into optical port.
CASSETTE
CLEANING
TAPE
DIRECTION
OF CONNECTOR
MOTION
ANGLED CONNECTOR
DETAIL DRAWING
PROTRUSIONS
PROTRUSIONS
14. Maintenance
The maintenance of the indexing terminal installation
includes cleaning :
14.1. HMFOC cleaning
14.1.1. Inspect the complete multi fiber ferrule on the
adapter side as well as on the connector side with a low
resolution microscope as specified in IEC 61300-3-35
with floodlight (side illumination). If dirt is detected, follow
cleaning instructions below.
14.1.2. CommScope recommends dry cleaning first
with a tool such as the US CONEC IBC brand cleaning
tool - MT Series (PN 15639). Follow the cleaning instruc-
tions included with the tool.
14.1.3. To clean the connector on the tail (female, plug,
non-pinned) place the blue plug adapter on the end of the
tool, align the keys of the tool and the connector, rotate
the thumb wheel clock wise two ‘clicks’.
14.1.4. To clean the adapter on the terminal (male,
jack, pinned) place the red plug adapter on the end of the
tool, align the keys of the tool (slot) and the adapter
(square notch), rotate the thumb wheel clock wise two
‘clicks’.
14.1.5. If dry cleaning does not produce the desired
results, use a small amount of solvent on one or more
Chemtronics ® Covenrty ™ 2.5mm fiber optic swabs.
A non isopropyl alcohol (IPA) solvent is recommended
(such as Chemtronix PX Wash). Using the swabs, re-
move contaminates/dirt from the ferrule end face. Always
use a new swab for each connector.
14.1.6. If there is a need to moisten a connector end
face, it must be blown dry with clean compressed air,
then dry clean the connector as detailed described in step
14.1.2 untill step 14.1.4.
14.1.7. After cleaning, inspect the multi fiber ferrule on
adapter and connector side again.