1  

Description

 

 

The cooling unit XRCA-5001-OA is designed to provide a continues flow of cooling 

 

 

fluid at a constant temperature.  

 

Oil circulates between the cooling unit and the heat source.  

 

The oil is chilled by an air-cooled refrigeration system (refrigerant R 134A).The  

 

cooling unit has a main switch to start the refrigeration system. The oil level in the  

 

reservoir is supervised by a level switch. If the level drops below minimum, the  

 

refrigeration cycle stops and the pilot lamp "Oil level low" is activated and the safety  

 

circuit is activated. 

 

The maximum cooling capacity of the cooling unit depends on the ambient and the  

 

fluid temperature. The cooling capacity is 5000W related to 40 °C ambient  

 

temperature and +30 °C ± 2°C fluid temperature  

 

Cooling hoses are connected to the cooling unit via screw connection (M26x1,5). Oil  

 

inlet and outlet are marked with symbols:  

 

Inlet:

O

Outlet:

O

 

 

 

The oil outlet temperature is monitored by an electronic temperature controller.  

 

 

This controller is provided with additional contacts and for temperature maximum  

 

signal. 

 

The signals "temperature maximum" and "oil level low" are combined in a potential  

 

free safety circuit. The oil pressure of the pump is limited by an adjustable bypass  

 

valve. 

 
 

Note: Flow switch, thermal switch and bypass valve are factory adjusted  

 

 

according to specification. 

 

Use only Shell Daial S4 ZX-I to refill the cooling circuit! 

Printed from SAP on 7.09.2015

SAP-DocNr: 50005555_001

Summary of Contents for XRCA-5001-OA

Page 1: ...engasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 5001 OA Doc No 50005555 Page 1 of 15 Industrial X Ray Chiller Manual XRCA 5001 OA For OEM Printed from SAP on 7 09 2015 SAP DocNr...

Page 2: ...y Version Date Author Amendment s Status 1 0 03 31 2010 M Schmid Layout Structure 07 23 2010 St Haferl Released 2 0 10 30 2012 R Moser Layout Structure 3 0 07 01 2013 R Moser Update Chapter 2 3 4 0 30...

Page 3: ...Run cooling unit always with correct coolant level otherwise damage of pump may occur Only use cooling hoses with sufficient pressure and coolant resistance Never operate damaged or leaking equipment...

Page 4: ...5 Page 4 of 15 Contents 1 Description 5 2 Technical data 6 3 Performance chart 7 4 Setting to work 8 5 Settings 9 6 Maintenance 11 7 Outline drawing 12 8 Circuit Diagrams 13 8 1 Electrical diagram 13...

Page 5: ...um cooling capacity of the cooling unit depends on the ambient and the fluid temperature The cooling capacity is 5000W related to 40 C ambient temperature and 30 C 2 C fluid temperature Cooling hoses...

Page 6: ...30 V available in 50 Hz or 60 Hz version Current consumption 8 2 A Noise level 70 dB A distance 1 m in any direction Airflow at 50Hz 60Hz 3400 m3 h 4400 m3 h Safety class IP 33 Environmental specifica...

Page 7: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 5001 OA Doc No 50005555 Page 7 of 15 3 Performance chart Printed from SAP on 7 09 2015 SAP DocNr 50005555_001...

Page 8: ...rake at caster remove side cover connect hoses with cooling unit remove plug of filler fill the cooling unit with oil up to top indicator establish electrical connection Note Operate equipment only wi...

Page 9: ...LED on cooling unit ready LED off Temperature below anti freezing running of refrigeration system impossible Oil outlet temperature rated value T1 Press SET key and keep it pressed rated value is sho...

Page 10: ...Hysterese of T3 adjusted at 2 5 C symmet rical to parameter P1 Example Value P1 32 5 C Temperature maximum T3 fault indication ON 35 0 C fault indication OFF 20 0 C Adjust new value with keys UP or D...

Page 11: ...er of the cooling unit clean Regularly once a week check this heat exchanger and if necessary clean it Clean fins of heat exchanger Remove dust by forcing it out in the opposite direction from which i...

Page 12: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 5001 OA Doc No 50005555 Page 12 of 15 7 Outline drawing Printed from SAP on 7 09 2015 SAP DocNr 50005555_001...

Page 13: ...Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 5001 OA Doc No 50005555 Page 13 of 15 8 Circuit Diagrams 8 1 Electrical diagram Printed from SAP on 7 09 2015 SAP DocNr 50005...

Page 14: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 5001 OA Doc No 50005555 Page 14 of 15 8 2 Block diagram Printed from SAP on 7 09 2015 SAP DocNr 50005555_001...

Page 15: ...OMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 5001 OA Doc No 50005555 Page 15 of 15 9 Declaration of conformity Printed from SAP on 7 09 2015 SAP DocNr 50005555_001...

Reviews: