The cooling unit XRCA-5001-OA is designed to provide a continues flow of cooling
fluid at a constant temperature.
Oil circulates between the cooling unit and the heat source.
The oil is chilled by an air-cooled refrigeration system (refrigerant R 134A).The
cooling unit has a main switch to start the refrigeration system. The oil level in the
reservoir is supervised by a level switch. If the level drops below minimum, the
refrigeration cycle stops and the pilot lamp "Oil level low" is activated and the safety
circuit is activated.
The maximum cooling capacity of the cooling unit depends on the ambient and the
fluid temperature. The cooling capacity is 5000W related to 40 °C ambient
temperature and +30 °C ± 2°C fluid temperature
Cooling hoses are connected to the cooling unit via screw connection (M26x1,5). Oil
inlet and outlet are marked with symbols:
Inlet:
O
↓
Outlet:
O
↑
The oil outlet temperature is monitored by an electronic temperature controller.
This controller is provided with additional contacts and for temperature maximum
signal.
The signals "temperature maximum" and "oil level low" are combined in a potential
free safety circuit. The oil pressure of the pump is limited by an adjustable bypass
valve.
Note: Flow switch, thermal switch and bypass valve are factory adjusted
according to specification.
Use only Shell Daial S4 ZX-I to refill the cooling circuit!