Temperature maximum T3
Press keys (UP) and (DOWN) simultaneously for
4 seconds, P1 is shown on display
Press SET-key and keep it pressed (value of P1 is
shown on display)
Note:
Hysterese of T3 adjusted at ± 2,5°C symmet-
rical to parameter P1.
Example:
Value P1:
32.5°C
•
Temperature maximum
T3:
•
fault indication ON:
35.0°C
•
fault indication OFF:
20.0°C
•
Adjust new value with keys (UP) or (DOWN)
•
Release SET-key (P1 is shown on display)
•
Press keys (UP) and (DOWN) simultaneously for 4
seconds to return to run-mode (if there is no changing of
parameters, the temperature controller returns automatically
into run-mode after 45 seconds).