Absolute humidity
Range:
0 to 400 g/m
3
Accuracy: ±1,5g/m
3
at air temperature T < 25°C (104 °F), for more details see
graphs below
Specific humidity
2
Accuracy:
±2g/kg at air temperature T < 35°C (95°F)
Range:
0 to 550 g/kg
Mixing ratio
2
Accuracy:
±2g/kg at air temperature T < 35°C (95°F)
Range:
0 to 995 g/kg
Specific enthalpy
2
Accuracy:
± 3kJ/kg at air temperature T < 25°C (77°F)
Range:
0 to 995 kJ/kg
3
The values computed from ambient temperature and relative humidity including their
accuracy you can exactly determine by the program
Conversions
. It is free to
download at
http://www.cometsystem.cz/dnload/conversions.exe
Response time
with stainless steel mesh sensor cover (F5200B) and bronze sensor cover
(F0000 - selectable option), air flow approximately 1 m/s:
temperature: T3319(L), T3419(L), T7311(L), T7411(L) t90 < 6 min (temperature
step 20 °C (36 °F) )
T3319P, T3419P t90 < 16 min (temperature step 20 °C (36 °F) )
relative humidity:
t90 < 30 s (humidity step 65 %RH, constant temperature)
atmospheric press.
4
: t90 <44s
Recommended calibration interval
: 1 year
2
This value depends on the barometric pressure. If device doesn’t support pressure measurement, then
constant value stored inside device memory is used. Default value preset by manufacturer is 1013hPa and
can be changed by user’s software.
3
This maximum is reached under conditions about 70°C/100%RH or 80°C/70%RH
4
There is possible to change response time. For more details see file “Description of calibration and
adjustment procedure.pdf”, which is installed together with Users software.