Setting of pressure correction for altitude 

For  devices  T7310  and  T7410  only!  If  there  is  necessary  to  read  value  of  atmospheric 

pressure corrected with respect to altitude, it is possible to set corresponding pressure offset for this 
altitude.  It  is  possible  to  do  it  with  Users  software.  This  value  is  then  automatically  added  to 
measured pressure value. The resultant value of pressure with added correction for altitude must be 
between  300  hPa  and  1350  hPa  (from  4.351  PSI  to  19.580  PSI).  Outside  this  range  device  sends 
value of Error 2! 

Description of communication protocols 

Detailed  description  of  each  communication  protocols  including  examples  of 

communication  is  available  in  individual  document  “

Description  of  communication  protocols  of 

Txxxx series”

 which is free to download a

www.cometsystem.cz

 

 
Note:  After switching  ON the power of the  device  it can last  up to  2 s before the  device  starts to 
communicate and measure! 

Modbus RTU 

Control units communicate on master-slave principle in half-duplex operation. Only master 

can  send  request  and  only  addressed  device  responds.  During  sending  of  request  no  other  slave 
station should respond. During communication, data transfer proceeds in binary format. Each Byte 
is sent as eight bit data word in format: 1 start bit, data word 8 bit (LSB first), 2 stop bits

1

, without 

parity. Device supports communication speed from 110Bd to 115200Bd. 
Sent request and response have syntax: ADDRESS OF DEVICE – FUNCTION – Modbus CRC 
 

Supported functions 

03 (0x03)

: Reading of 16-bit registers (Read Holding Registers) 

04 (0x04)

: Reading of 16-bit input gates (Read Input Registers) 

16 (0x10)

: Setting of more 16-bit registers (Write Multiple Registers) 

 

Jumper and button 

Jumper  and  button  are  located  next  to  connection  terminals.  If  communication  protocol 

Modbus is selected the function of jumper and button is as follows: 

 

Jumper  opened  –  device  memory  is  protected  from  writing,  from  device  side  it  is  only 
enabled  to  read  measured  value,  writing  to  memory  is  disabled  (no  change  of  device 
address, communication speed and LCD setting is enabled). 

 

Jumper closed – writing to device memory is enabled by means of User’s software. 

 

Jumper  opened  and  button  shortly  pressed  –  device  goes  to  Info  mode,  see  chapter  „Info 
mode“

 

Jumper  closed  and  button  pressed  for  longer  than  six  seconds  –  causes  restoring  of 
manufacturer  setting 

of  communication  protocol

,  i.e.  sets  Modbus  RTU  communication 

protocol, device address sets to 01h and communication speed to 9600Bd - after button press 
there  is  “dEF”  message  blinking  at  LCD  display.  Six  seconds  later  message  “dEF”  stays 
shown, it means manufacturer setting 

of communication protocol

 is done. 

                                                 

1

 Device sends two stop bits, for receive one stop bit is enough. 

Summary of Contents for T3311

Page 1: ...lative humidity and other derived humidity values T3311 T3313 T3411 Programmable transmitter of atmospheric pressure temperature relative humidity and other derived humidity values T7310 T7410 with RS...

Page 2: ...shown on dual line LCD display If there are two values displayed at one LCD line they are periodically switched between both readings with period of 4 seconds Display can be switched OFF totally too...

Page 3: ...ices with RS485 output there is recommended to use shielded twisted copper cable maximal length 1200m The cable must be located at indoor rooms Nominal cable impedance should be 100 loop resistance ma...

Page 4: ...4 IE SNC T33 4 11 T73 4 10 07 Dimensions T3311 T7310 Transmitter with RS232 Dimensions T3313 Transmitter with RS232...

Page 5: ...IE SNC T33 4 11 T73 4 10 07 5 Dimensions T3411 T7410 Transmitter with RS485 Typical application wiring connection of terminals Power from communication port Power from external voltage source...

Page 6: ...6 IE SNC T33 4 11 T73 4 10 07 Dimensions T3311L T3411L T7310L T7410L Connection see Appendix B Dimensions T3313L Connection see Appendix B...

Page 7: ...ght bit data word in format 1 start bit data word 8 bit LSB first 2 stop bits1 without parity Device supports communication speed from 110Bd to 115200Bd Sent request and response have syntax ADDRESS O...

Page 8: ...inH2O Int 10 kPa Int 100 Since FW version 02 44 there is possibility to read the all computed values from registers Variable Unit Address hex X Address dec X Format Size Status Dew point temperature...

Page 9: ...ess and check sum is reset in accordance with values stored in the device Communication speed and check sum are possible to change only if jumper is closed Jumper closed and button pressed for longer...

Page 10: ...g procedure to device s memory occurred or if damage of calibration data appeared At this state device does not measure and calculate values It is a serious error contact distributor of the device to...

Page 11: ...state of value If selected pressure unit is mBar or oz in2 or mmHg or inH2O or kPa there is shown only value number without corresponding pressure unit C F DP reading next to this symbol is calculate...

Page 12: ...Pa Range 600 8 70 450 0 17 72 240 9 139 2 60 00 1100 15 95 825 1 32 48 441 6 255 3 110 00 accuracy T 23 C 1 3 0 02 1 0 0 04 0 5 0 3 0 13 0 T 40 C 1 5 0 02 1 1 0 04 0 6 0 3 0 15 Else 2 0 0 03 1 5 0 06...

Page 13: ...ve humidity including their accuracy you can exactly determine by the program Conversions It is free to download at http www cometsystem cz dnload conversions exe Response time air flow approximately...

Page 14: ...ation It is not allowed to operate the device under conditions other than specified in technical parameters Devices are not designed for locations with chemically aggressive environment Temperature an...

Page 15: ...cted to the bus by connecting the corresponding pins of CANON connector for more information see the operation manual for ELO E214 The ELO E06D converter is an optional accessory for connection of tra...

Page 16: ...16 IE SNC T33 4 11 T73 4 10 07 Appendix B Connection of transmitters version TxxxxL...

Reviews: