Comet T3311 Instruction Manual Download Page 11

IE-SNC-T33(4)11+T73(4)10-07 

 

  11 

 

Computed value – calculation of computed value is not possible (error during measurement 
of temperature or relative humidity). 

Error 3 

-  there  is  a  reading  „Err3“  on  LCD  display  upper  line.  Error  of  internal  A/D  converter 

appeared  (converter  does  not  respond,  probably  damage  of  A/D  converter).  This  error  does 
NOT affect pressure measurement. Rest values are NOT measured. It is a serious error, contact 
distributor of the device. 

Error 4 

-  there  is  a  reading  „Err4“  on  LCD  display.  It  is  internal  device  error  during  pressure 

sensor  initialization.  Under  this  condition  device  does  NOT  measure  atmospheric  pressure. 
Value  read  from  device  is  -999.9.  Pressure  sensor  is  probably  damaged.  It  is  a  serious  error, 
contact distributor of the device. 

 

Readings on LCD display 

°C, °F - 

reading next to this symbol is measured temperature or error state of value. 

 

%RH  - 

reading next to this symbol is measured relative humidity or error state of value. 

 

hPa,  PSI,  inHg,    - 

reading  next  to  this  symbol  is  measured  pressure  or  error  state  of  value.  If 

selected pressure unit is 

mBar 

or

 oz/in

2

 or

 mmHg

 or

 inH

2

O

 or 

kPa

, there is shown only 

value (number) without corresponding pressure unit! 

 

°C / °F  DP  - 

reading  next  to  this  symbol  is  calculated  dew  point  temperature  or  error  state  of 

value. 

 

g/m

3

  -  

reading next to this symbol is calculated absolute humidity or error state of value. 

 

g/kg  - 

reading  next  to  this  symbol  is  calculated  specific  humidity  or  mixing  ratio  (depends  on 

device setting) or error state of value. 

 

If specific enthalpy is selected, there is shown only value (number) without corresponding unit! 

 

3  -  

this symbol is on if jumper is closed. 

 
 

Technical parameters of the device: 

RS 485 Interface: 

Receiver-Input Resistance:  96 kΩ 
Devices on bus: 

 

max. 256  (1/8 Unit Receiver Load) 

 

Measuring parameters

Ambient temperature

 

T3311(L), T3411(L), T7310(L), T7410(L): 

Measuring range: 

-30 to +80 °C 

Display resolution: 

0.1 °C 

Accuracy: 

 

± 0.4 °C 

Ambient temperature

 

T3313(L): 

Measuring range: 

-30 to +125 °C 

Display resolution: 

0.1 °C 

Accuracy: 

 

± 0.4 °C from -30 to +100 °C, otherwise 0,4% fro reading 

Relative humidity

 (reading is temperature compensated at entire temperature range): 

Measuring range:  

0 to 100 %RH (see Device installation) 

Display resolution:   0.1 %RH 
Accuracy:  

 

± 2.5 %RH from 5 to 95 %RH at 23 °C 

Summary of Contents for T3311

Page 1: ...lative humidity and other derived humidity values T3311 T3313 T3411 Programmable transmitter of atmospheric pressure temperature relative humidity and other derived humidity values T7310 T7410 with RS...

Page 2: ...shown on dual line LCD display If there are two values displayed at one LCD line they are periodically switched between both readings with period of 4 seconds Display can be switched OFF totally too...

Page 3: ...ices with RS485 output there is recommended to use shielded twisted copper cable maximal length 1200m The cable must be located at indoor rooms Nominal cable impedance should be 100 loop resistance ma...

Page 4: ...4 IE SNC T33 4 11 T73 4 10 07 Dimensions T3311 T7310 Transmitter with RS232 Dimensions T3313 Transmitter with RS232...

Page 5: ...IE SNC T33 4 11 T73 4 10 07 5 Dimensions T3411 T7410 Transmitter with RS485 Typical application wiring connection of terminals Power from communication port Power from external voltage source...

Page 6: ...6 IE SNC T33 4 11 T73 4 10 07 Dimensions T3311L T3411L T7310L T7410L Connection see Appendix B Dimensions T3313L Connection see Appendix B...

Page 7: ...ght bit data word in format 1 start bit data word 8 bit LSB first 2 stop bits1 without parity Device supports communication speed from 110Bd to 115200Bd Sent request and response have syntax ADDRESS O...

Page 8: ...inH2O Int 10 kPa Int 100 Since FW version 02 44 there is possibility to read the all computed values from registers Variable Unit Address hex X Address dec X Format Size Status Dew point temperature...

Page 9: ...ess and check sum is reset in accordance with values stored in the device Communication speed and check sum are possible to change only if jumper is closed Jumper closed and button pressed for longer...

Page 10: ...g procedure to device s memory occurred or if damage of calibration data appeared At this state device does not measure and calculate values It is a serious error contact distributor of the device to...

Page 11: ...state of value If selected pressure unit is mBar or oz in2 or mmHg or inH2O or kPa there is shown only value number without corresponding pressure unit C F DP reading next to this symbol is calculate...

Page 12: ...Pa Range 600 8 70 450 0 17 72 240 9 139 2 60 00 1100 15 95 825 1 32 48 441 6 255 3 110 00 accuracy T 23 C 1 3 0 02 1 0 0 04 0 5 0 3 0 13 0 T 40 C 1 5 0 02 1 1 0 04 0 6 0 3 0 15 Else 2 0 0 03 1 5 0 06...

Page 13: ...ve humidity including their accuracy you can exactly determine by the program Conversions It is free to download at http www cometsystem cz dnload conversions exe Response time air flow approximately...

Page 14: ...ation It is not allowed to operate the device under conditions other than specified in technical parameters Devices are not designed for locations with chemically aggressive environment Temperature an...

Page 15: ...cted to the bus by connecting the corresponding pins of CANON connector for more information see the operation manual for ELO E214 The ELO E06D converter is an optional accessory for connection of tra...

Page 16: ...16 IE SNC T33 4 11 T73 4 10 07 Appendix B Connection of transmitters version TxxxxL...

Reviews: