Comet SP015 Quick Manual Download Page 4

 
  

- 4 - 

BATTERY REPLACEMENT 

If  it  is  necessary  to  replace  the  batteries  in  the 
Batteries holder directly at the installation site, first 
unscrew  the  transducer  from  the  holder  posts  (2 
M4 screws) and move it to the right of the Batteries 
holder. Temporarily attach the transducer with only 
one screw to the right post of the Batteries holder 
as  shown.  Do  not  let  the  transducer  hang  by  the 
power cord. Remove the four screws at the corners 
of case the Batteries holder and remove the cover. 
Replace  the  batteries  with  new  ones  and 
reassemble all the assembly in reverse. 
 

TECHNICAL PARAMETERS 

 

recommended batteries:   Tadiran SL2770/S, lithium, 3.6 V, size "C" 

 

output voltage:  

 

3.6 V DC 

 

continuous output current:  max. 200 mA 

 

short circuit protection of output: no 

(recommended batteries are not dangerous in the event of a short circuit)

 

 

power output:  

 

connection adapter for transducers Wx8xx 

 

working position:    

vertically, cable gland down 

 

IP class:   

 

IP65  (after mounting the connection adapter to the Wx8xx sensor) 

 

operating temperature:   -30 to +60 °C 

 

weight without batteries:  200 g (including connection adapter) 

 

maximum dimensions: 

120 x 93 x 43 mm 

 
 
 

DEVICE DIMENSIONS  

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 