IE-LOG-S0841-07 

 

  5 

 

optionally switch ON displaying of MIN/MAX values on the LCD 

 

reset memory of MIN/MAX values (if needed) 

 

check free space in data memory, optionally erase data memory of the logger 

 

enter password if protection against unauthorized manipulation with the logger is necessary  

 
Logging interval between to subsequent measurements is specified by the user. Memorizing of the 
first value is synchronized with the internal real time clock, so the logging is performed at sharp 
multiples of minutes, hours and days. E.g. after starting the logging with the 15 minute interval the 
first value is not stored immediately, but after the internal clock gets the status of a quarter, a half or 
a whole hour. After starting the logging with the 6 hours interval the first value is stored at that 
whole hour to perform the storing also at 00.00, i.e. at the beginning of the day. First storing is 
performed at 6.00,12.00, 18.00 or 00.00hour - at the hour from the above nearest to the logging 
start. After communication with computer or after start by magnet logger automatically waits for the 
nearest whole multiple of time and then first measurement is performed. This is also necessary to 
take into account when setting time of automatic logger switch ON. 

Notice:

 If logger operates as permanently connected to the computer, using of magnet start/stop 

is disabled. 

 To  enable  the  logger  control  by  magnet  is  suitable  only  in  cases,  when  possibility  of 

unauthorized manipulation to the logger operation is eliminated. 

 

Reading on the display in usual operation (logger switched ON) 

 
After switching ON the logger all LCD symbols are displayed for checking 
the display. 
 
Then actual date and time in logger is displayed for approximately 4 s. 
 
 
Consequently  reading  of  estimated  remaining  battery  capacity  for 
approximately  2  s  is  displayed  (values  0  to  100%).  It  is  valid  if  logger  is 
operated at temperature from -5 to +35°C. If logger is operated often outside 
of  the  above  temperature  range  battery  life  can  be  reduced  to  75%,  i.e.  if 
displayed  remaining  battery  capacity  drops  below  25%,  it  is  recommended 
to replace the battery. 
 

If display is switched ON

, actual reading of measured values is displayed - 

channel 1 temperature (°C 1) at the LCD upper line, channel 2 temperature 
(°C  2)  at  the  LCD  lower  line.  Symbol  LOG  indicates  data  logging  in 
progress – if it blinks, data memory is filled up to more than 90%. 
 
Other measured or calculated values are displayed always after 5 seconds. 
Instrument  now  displays  channel  1  temperature  (°C  1)  and  temperature 
difference between channels (lower line "DIF"). T

DIF

 = T1 – T2 

 
Instrument  now  displays  actual  binary  state  of  channel  3  (LCD  upper  line) 
and  channel  4  (LCD  lower  line).  Symbol  On  indicates  closed  contact, 
symbol OFF indicates open (or not connected) contact. 
 

Summary of Contents for S0841

Page 1: ...LOGGER S0841 Dual channel temperature logger with two binary inputs and LCD display Instruction Manual ...

Page 2: ...e when alarm is indicated permanently till alarm memory reset Alarm function is enabled or disabled for each channel individually Binary channels have no alarm modes Logging mode can be adjusted as non cyclic when logging stops after filling the memory In cyclic mode oldest stored values are overwritten by new In addition logging mode can be selected when logging is active only if measured value i...

Page 3: ...lock adjustable from computer integrated calendar including leap years Error of internal RTC 200 ppm i e 0 02 17 28 s in 24 h Power Lithium battery 3 6 V size AA Battery life typical data download to PC is carried out about once a week 5 years in continuous on line mode with interval 1 minute approximately 3 years in continuous on line mode with interval 10 seconds approximately 1 year Notice The ...

Page 4: ...heck or optionally set the logger real time clock select suitable logging interval it has no influence to binary inputs record select logging mode cyclic or non cyclic set for each channel type of temperature probe Pt1000 or Ni1000 disable channels for logging with not connected temperature probe switch ON the logger or switch OFF if it is about to be switched ON by magnet or automatically with de...

Page 5: ... permanently connected to the computer using of magnet start stop is disabled To enable the logger control by magnet is suitable only in cases when possibility of unauthorized manipulation to the logger operation is eliminated Reading on the display in usual operation logger switched ON After switching ON the logger all LCD symbols are displayed for checking the display Then actual date and time i...

Page 6: ...ing therefore does not run Displayed symbol indicates logger is ON If information on actual measured values is needed it is possible anytime to display reading display by means of the magnet only if communication adapter is not connected permanently Plug magnet into guide slots from logger front side for approximately 4 s and wait till reading on the display appears f logger has enabled the functi...

Page 7: ... or battery disconnection for longer period than approximately 30 s after display switch ON during date and time display all four arrows appears as warning to check or set it again from the computer However all logger functions work without limitation If reading BAT is displayed periodically on LCD upper line for 1 s with 10 s interval the end of estimated battery life is coming however logger fun...

Page 8: ...ry to enable the function from the PC software before Notice It is not possible to combine this function with function of switching logger OFF by magnet User software enables to select only one of them or none Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD upper line After decimal point appearance it is necessary immediately till...

Page 9: ...ery replacement This step is necessary to evaluate properly the remaining battery capacity Old battery or logger itself after its life is necessary to liquidate ecologically Connection of female ELKA 3008V connector for temperature probe and binary contact It is possible to order as an optional accessory connector for connection of temperature probe Signal cable should be shielded round cross sect...