Replacement procedure:
•
switch off the logger by the PC program or by magnet (if low battery allows)
•
unscrew four corner screws and remove the lid
•
remove old battery by pulling the glued tape
•
insert new battery
respecting the correct polarity
(see s and – near battery holder). If
you connect new battery up to 30 s, all logger settings remain unchanged. In the opposite case
check by means of PC program all settings, especially the real time clock in logger.
Attention,
inserted battery with incorrect polarity causes logger damage!
•
put the lid back again and screw four screws. Be careful to have the rubber sealing properly in
the slot and tighten the screws properly to ensure the water resistance of the instrument.
connect logger to the computer and
write to it the information on battery replacement
(menu Configuration/Battery replacement). This step is necessary to evaluate properly the
remaining battery capacity.
Old battery or logger itself (after its life) is necessary to liquidate ecologically!
Connection of female connector ELKA 3008V to temperature probe
It is possible to order as an optional accessory connector for
connection of temperature probe. Probe cable should be shielded, round
cross section of minimum diameter of 3.5 mm and maximum diameter
of 5 mm, maximum length of 10 m and as large as possible cross
section of cords (stranded wires). Outer diameter must be from 1 to
1.55 mm including insulation. Connect connector to temperature probe
accordingly to the figure by means of the manual included in delivery.
Choose the most suitable plastic insert accordingly to the diameter of
stranded wires of the cable. The IP67 protection is guaranteed only if
connector is assembled correctly!
Instruments passed the following electromagnetic compatibility (EMC) tests:
radiation: EN 55022
class B
immunity: EN 61000-4-2
(levels 4/8 kV, class A)
EN 61000-4-3
(intensity of electromagnetic field 3 V/m, class A)
EN 61000-4-4
only types with external probes (levels 1/0,5 kV, class A)
EN 61000-4-6
only types with external probes (intensity of electromagnetic
field 3 V/m, class A)
8