Replacement procedure: 

 

switch off the logger by the PC program or by magnet (if low battery allows) 

 

unscrew four corner screws and remove the lid 

 

remove old battery by pulling the glued tape 

 

insert new battery 

respecting the correct polarity

 (see s and – near battery holder). If 

you connect new battery up to 30 s, all logger settings remain unchanged. In the opposite case 
check by means of PC program all settings, especially the real time clock in logger. 

Attention, 

inserted battery with incorrect polarity causes logger damage!

 

 

put the lid back again and screw four screws. Be careful to have the rubber sealing properly in 
the slot and tighten the screws properly to ensure the water resistance of the instrument. 

connect logger to the computer and 

write to it the information on battery replacement 

(menu Configuration/Battery replacement). This step is necessary to evaluate properly the 
remaining battery capacity. 

 

Old battery or logger itself (after its life) is necessary to liquidate ecologically! 

 
 

 

Connection of female connector ELKA 3008V to temperature probe 

 

It is possible to order as an optional accessory connector for 

connection of temperature probe. Probe cable should be shielded, round 
cross section of minimum diameter of 3.5 mm and maximum diameter 
of 5 mm, maximum length of 10 m and as large as possible cross 
section of cords (stranded wires). Outer diameter must be from 1 to 
1.55 mm including insulation. Connect connector to temperature probe 
accordingly to the figure by means of the manual included in delivery. 
Choose the most suitable plastic insert accordingly to the diameter of 
stranded wires of the cable. The IP67 protection is guaranteed only if 
connector is assembled correctly! 

 

 
 

Instruments passed the following electromagnetic compatibility (EMC) tests: 

radiation:  EN 55022 

 

class B  

immunity:  EN 61000-4-2  

(levels 4/8 kV, class A) 

 

 

EN 61000-4-3  

(intensity of electromagnetic field 3 V/m, class A) 

 

 

EN 61000-4-4  

only types with external probes (levels 1/0,5 kV, class A) 

 

 

EN 61000-4-6  

only types with external probes (intensity of electromagnetic 

 

 

 

 

 

 

field 3 V/m, class A) 

8  

IE-LOG-S0141-04 

Summary of Contents for LOGGER S0141

Page 1: ...LOGGER S0141 Four channel temperature logger with display Instruction Manual...

Page 2: ...written by new In addition logging mode can be selected when logging is active only if measured value is out of adjusted alarm limits Stored values can be transferred from logger memory to the PC by m...

Page 3: ...tion by self adhesive Dual Lock applied to clean flat surface Not allowed manipulation it is not allowed to operate instrument with not connected temperature probes if all probes are not connected unc...

Page 4: ...f the logger enter password if protection against unauthorized manipulation with the logger is necessary Logging interval between to subsequent measurements is specified by the user Memorizing of the...

Page 5: ...al temperature Notice if alarm is active no MIN MAX reading is displayed If display is switched OFF all above readings are displayed up to estimated remaining battery capacity and then display goes ou...

Page 6: ...New initialization of logger can occur in switching ON the logger immediately after displaying of all LCD segments for checking e g after replacement of totally discharged battery for new one State i...

Page 7: ...magnet User software enables to select only one of them or none Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD upper line After dec...

Page 8: ...after its life is necessary to liquidate ecologically Connection of female connector ELKA 3008V to temperature probe It is possible to order as an optional accessory connector for connection of temper...

Reviews: