Comet LOGGER S0141 Instruction Manual Download Page 6

 

 

Alarm indication on the display 

It is necessary to enable alarm function from the PC and set for each quantity lower and upper 

limit. If measured value is inside of set limits, alarm of proper quantity is not active. If value of 
measured quantity gets outside of set limits, alarm of proper quantity is active and it is indicated on 
the display. It is possible to select "memory alarm mode" when alarm is indicated permanently up 
to reset from the PC. 

 
Active alarm is indicated (if display is ON) by blinking of the value of 
proper quantity on the display and arrow symbol appears at the LCD upper 
side at the same time. Arrow 1 indicates active alarm for channel 1 , arrow 2 
for channel 2, arrow 3 for channel 3 and arrow 4 for channel 4 temperature. 

Notice:

 if logger is operated at low temperature (below approximately 

-5 °C), alarm indication by blinking can be indistinct. Indication by arrows 
works correctly. 
 

 

Messages displayed on the LCD beyond usual operation 

If measured value is out of the measurable or displayable range numeric 
reading is replaced by hyphens. 
If memory is totally full filled in non-cyclic logging mode, logger is 
switched OFF and message MEMO FULL appears on the LCD. It appears 
also if logger is operated with switched OFF display. 
 
New initialization of logger can occur in switching ON the logger 
(immediately after displaying of all LCD segments for checking) e.g. after 
replacement of totally discharged battery for new one. State is indicated by 
INIT reading. It can be displayed for about 12 s. 
If battery voltage drop occurred since last internal clock setting below 
critical limit or battery disconnection for longer period than approximately 
30 s, after display switch ON (during date and time display) all four arrows 
appears as warning to check or set it again from the computer. However all 
logger functions work without limitation. 
If reading BAT is displayed periodically on LCD upper line (for 1 s with 10 
s interval), the end of estimated battery life is coming – however logger 
functions are not limited. Replace battery as soon as possible! 
If reading BAT is displayed permanently, battery voltage is low and logger 
is  not possible to switch ON. If logger was switched ON before it, data 
logging is stopped and logger is switched OFF. Communication 
with computer can temporarily work. Replace battery as soon as possible! 
 

 

Start / stop by magnet 

The function must be enabled from the PC before. If only switching OFF by magnet is enabled, 

it is of course necessary to switch ON the logger from the computer. 

Notice: 

It is not possible to combine function switching OFF by magnet and MIN/MAX 

memory clear by magnet! User software enables only one of them to select. 

 

6  

IE-LOG-S0141-04 

Summary of Contents for LOGGER S0141

Page 1: ...LOGGER S0141 Four channel temperature logger with display Instruction Manual...

Page 2: ...written by new In addition logging mode can be selected when logging is active only if measured value is out of adjusted alarm limits Stored values can be transferred from logger memory to the PC by m...

Page 3: ...tion by self adhesive Dual Lock applied to clean flat surface Not allowed manipulation it is not allowed to operate instrument with not connected temperature probes if all probes are not connected unc...

Page 4: ...f the logger enter password if protection against unauthorized manipulation with the logger is necessary Logging interval between to subsequent measurements is specified by the user Memorizing of the...

Page 5: ...al temperature Notice if alarm is active no MIN MAX reading is displayed If display is switched OFF all above readings are displayed up to estimated remaining battery capacity and then display goes ou...

Page 6: ...New initialization of logger can occur in switching ON the logger immediately after displaying of all LCD segments for checking e g after replacement of totally discharged battery for new one State i...

Page 7: ...magnet User software enables to select only one of them or none Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD upper line After dec...

Page 8: ...after its life is necessary to liquidate ecologically Connection of female connector ELKA 3008V to temperature probe It is possible to order as an optional accessory connector for connection of temper...

Reviews: