Settings
Note:
The temperature controller and the bypass valve are adjusted according specifica-
tion. Settings of water outlet temperature, temperature maximum (parameter P1),
anti freezing, and bypass valve should not be changed. If it is necessary to change
these settings refer to the following:
Set maximum pressure:
Basic adjustment < 6,7 bar
Solve sealing
Increasing maxium pressure:
Turn set screw clockwise
reducing maximum pressure: turn
set screw anti-clockwise
Set flow switch
Unscrew fixing screw of contactor
head
Scale up +: move contactor head in
the direction of the oil flow
Scale down -: move contactor head
against the direction of the oil flow
Screw on fixing screw of contactor
head