Industrial X-Ray 

 

Chiller Manual XRCA-3001-WA 

For OEM 

 

Summary of Contents for 10008643

Page 1: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 3001 WA Doc No 50005554 Page 1 of 16 Industrial X Ray Chiller Manual XRCA 3001 WA For OEM...

Page 2: ...WA Doc No 50005554 Page 2 of 16 Document history Version Date Author Amendment s Status 1 0 03 31 2010 M Schmid Layout structure 07 23 2010 St Haferl Released 2 0 10 30 2012 R Moser Layout structure...

Page 3: ...y 230 V 50 60 Hz 208 V 50 60 Hz Run cooling unit always with correct coolant level and clean filter regular oth erwise damage of pump may occur Only use cooling hoses with sufficient pressure and cool...

Page 4: ...m XRCA 3001 WA Doc No 50005554 Page 4 of 16 Contents 1 Description 5 2 Technical data 6 3 Performance chart 7 4 Setting to work 8 5 Settings 9 6 Maintenance 12 7 Outline Drawing 13 8 Circuit Diagrams...

Page 5: ...acity of the cooling unit depends on the ambient and the fluid temperature The cooling capacity is 3000W related to 40 C ambient temperature and 25 C 2 C fluid temperature Cooling hoses are connected...

Page 6: ...ge rating 230 V 50 60 Hz switchable to 208 V 50 60 Hz Current consumption 9 3 A Noise level 65 dB A distance 1 m in any direction Airflow at 50Hz 60Hz 1700 m3 h 2200 m3 h Safety class IP 33 Environmen...

Page 7: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 3001 WA Doc No 50005554 Page 7 of 16 3 Performance chart...

Page 8: ...r about 3 cm below thread For ambient temperatures lower than 10 C use anti freeze We recommend Ethylenglycol with tap water in a mix of 2 3 check coolant level and refill if necessary Establish elect...

Page 9: ...ng and bypass valve should not be changed If it is necessary to change these settings refer to the following Set maximum pressure Basic adjustment 6 7 bar Solve sealing Increasing maxium pressure Turn...

Page 10: ...Impossible Water outlet temperature rated value T1 Press SET key and keep it pressed rated value is shown on display Release SET key actual value is shown on display Note Hysteresis of T3 is adjusted...

Page 11: ...and keep it pressed value of P1 is shown on display Note Hysterese of T3 adjusted at 2 5 C symmet rical to parameter P1 Example Value P1 32 5 C Temperature maximum T3 fault indication ON 35 0 C fault...

Page 12: ...freeze About every 3 months check the filter of the pump for clean condition if necessary more often For this proceed as following Note If the filter is not in clean condition damage of pump and motor...

Page 13: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 3001 WA Doc No 50005554 Page 13 of 16 7 Outline Drawing...

Page 14: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 3001 WA Doc No 50005554 Page 14 of 16 8 Circuit Diagrams Electrical diagram...

Page 15: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 3001 WA Doc No 50005554 Page 15 of 16 Block diagram...

Page 16: ...COMET AG Herrengasse 10 CH 3175 Flamatt Tel 41 31 744 9000 www comet xray com XRCA 3001 WA Doc No 50005554 Page 16 of 16 9 Declaration of conformity...