Error 2 

There is a reading „Err2“ on LCD display. Pressure measurement error occurred. Value read from 
the device is -999.9. 

 

This  state  occur if measured pressure value 

with added correction for altitude

  is  outside 

of range  from  300 hPa  to  1350 hPa  (from  4.351  PSI  to  19.580  PSI)  or  the  pressure  sensor 
is damaged. Please check setting of pressure correction for altitude with User’s software. 

Error 4 

There  is  a  reading  „Err4“  on  LCD  display.  It  is  internal  device  error  during  pressure  sensor 
initialization.  Under  this  condition  device  does  NOT  measure  atmospheric  pressure.  Value  read 
from device is -999.9. Pressure sensor is probably damaged. It is a serious error, contact distributor 
of the device. 
 

Readings on LCD display 

hPa, PSI, inHg,  

Reading next to this symbol is measured pressure or error state of value. If selected pressure unit is 

mBar 

or

  oz/in

2

  or

  mmHg

  or

  inH

2

O

  or 

kPa

,  there  is  shown  only  value  (number)  without 

corresponding pressure unit! 

 

symbol

 3 

near by left display margin

 

Is on if jumper is closed. 
 

Technical parameters of the device: 

RS 485 Interface: 

Receiver-Input Resistance:  96 kΩ 

Devices on bus: 

 

max. 256  (1/8 Unit Receiver Load) 

 

Measuring parameters: 

Atmospheric pressure 

(T5410 only)

:

 

 

Unit 

hPa, mBar 

PSI 

mmHg 

inHg 

inH

2

oz/in

2

 

kPa 

Range 

600 

8.70 

450.0 

17.72 

240.9 

139.2 

60.00 

1100 

15.95 

825.1 

32.48 

441.6 

255.3 

110.00 

accuracy 

 

T=23 °C 

±1.3 

±0.02 

±1.0 

±0.04 

±0.5 

±0.3 

±0.13 

0≤T≤40°C 

±1.5 

±0.02 

±1.1 

±0.04 

±0.6 

±0.3 

±0.15 

Else 

±2.0 

±0.03 

±1.5 

±0.06 

±0.8 

±0.5 

±0.20 

 
Response time

 atmospheric pressure

2

: t90 <44s  

Recommended calibration interval:

  1 year 

Measuring interval and LCD display refresh:

 2 s 

Power:

 9 to 30 V dc 

 

2

  There  is  possible  to  change  response  time.  For  more  details  see  file  “Description  of  calibration 

and adjustment procedure.pdf”, which is installed together with Users software. 

Summary of Contents for T2314

Page 1: ...cturer reserves the right to make technical changes to the device without previous notice Misprints reserved Manufacturer is not responsible for damages caused by using the device in conflict with thi...

Page 2: ...www cometsystem com It supports make the adjustment of the device too This procedure is described at file Calibration manual pdf which is installed commonly with the software Change of some parameter...

Page 3: ...r side The cable should be led in one line i e NOT to tree or star Termination resistor should be located at the end For short distance other topology is allowed Terminate the network by a termination...

Page 4: ...4 IE SNC T23 4 14 04 Dimensions T2314 Transmitter with RS232 Dimensions T2414 Transmitter with RS485 Dimensions T2314L T2414L Connection see Appendix B...

Page 5: ...IE SNC T23 4 14 04 5 Typical application wiring connection of terminals T2314 power from communication port T2314 power from external voltage source T2414 RS485...

Page 6: ...bit registers Read Holding Registers 04 0x04 Reading of 16 bit input gates Read Input Registers 16 0x10 Setting of more 16 bit registers Write Multiple Registers Jumper and button Jumper and button ar...

Page 7: ...g Note In case there is a need for reading of measured values from the device with higher resolution than one decimal measured values in device are stored also in Float format which is not directly co...

Page 8: ...ce with setup software TSensor to communication protocol HWg Poseidon and set correct device address This communication protocol supports read temperature at C relative humidity one of computed value...

Page 9: ...sure or error state of value If selected pressure unit is mBar or oz in2 or mmHg or inH2O or kPa there is shown only value number without corresponding pressure unit symbol 3 near by left display marg...

Page 10: ...allowed manipulation It is not allowed to operate the device under conditions other than specified in technical parameters Devices are not designed for locations with chemically aggressive environmen...

Page 11: ...connector marked RS232 directly to the PC connect power to connector marked RS485 Power voltage 6V DC from an external acdc adapter connect to pin 9 0V connect to pin 5 Also mutually connect pin 2 and...

Reviews: