IE-SNC-T23(4)14-04
Instruction Manual
Device type
Pressure Output
Galvanic isolated
output
T2314
✓
RS232
-
T2414
✓
RS485
✓
Device is designed for measurement of atmospheric pressure at hPa, PSI, inHg, mBar,
oz/in
2
, mmHg, inH
2
O and kPa. It is possible to set pressure correction for altitude (offset). Device
setting is provided by user’s software, see latter. Device T2314 communicate via link RS232,
device T2414 via link RS485. Supported communication protocols are Modbus RTU, protocol
compatible with standard Advantech-ADAM, ARION and communication with HWg–Poseidon
devices. Devices are preset from manufacturer to Modbus RTU communication protocol. If you
would like to use different communication protocol then Modbus RTU, it is necessary preset them
through user’s software – see latter. Pressure sensor is placed inside the electronic case. Measured
value is optionally shown on LCD display. Display can be switched OFF too.
Use user’s software TSensor for setting of all device parameters (recommended).
It is free to download at
. It supports make the adjustment of the device
too. This procedure is described at file „
Calibration manual.pdf
“ which is installed commonly
with the software. Change of some parameters is possible to do without user’s software
with Windows hyperterminal (change of communication protocol, its parameters, LCD display
setting). It is described in file “
Description of communication protocols of Txxxx series
” which is
free to download at the same address.
Transmitter version T
xxxx
L
with watertight male connector instead of a cable gland
(RS232) or glands (RS485) is designed for easy connection/disconnection of communication cable.
Male Lumberg connector RSFM4 has protection IP67.
Models marked T
xxxx
Z
are non-standard versions of the transmitters. Description is not
included in this manual.
Please read instruction manual before the first device connection.
Device setting from the manufacturer
If special setting was not required in the order device is set from the manufacturer to the following
parameters:
communication protocol:
Modbus RTU
device address:
01H
communication speed:
9600Bd, without parity, 2 stop bits
display:
switched ON
pressure unit:
hPa
correction for altitude:
0 hPa
(absolute pressure)