Instruction Manual 

 
 

Device type 

Pressure  Output 

Galvanic isolated 

output  

T2314 

 

RS232 

T2414 

 

RS485 

 

 

Device  is  designed  for  measurement  of  atmospheric  pressure  at  hPa,  PSI,  inHg,  mBar, 

oz/in

2

, mmHg, inH

2

O and kPa. It is possible to set pressure correction for altitude (offset). Device 

setting  is  provided  by  user’s  software,  see  latter.  Device  T2314  communicate  via  link  RS232, 
device  T2414  via  link RS485.  Supported  communication  protocols  are  Modbus  RTU,  protocol 
compatible  with  standard  Advantech-ADAM,  ARION  and  communication  with  HWg–Poseidon 
devices.  Devices  are  preset  from  manufacturer  to  Modbus  RTU  communication  protocol.  If  you 
would like to use different communication protocol then Modbus RTU, it is necessary preset them 
through user’s software – see latter. Pressure sensor is placed inside the electronic case. Measured 
value is optionally shown on LCD display. Display can be switched OFF too. 

 

Use  user’s  software  TSensor  for  setting  of  all  device  parameters  (recommended). 

It is free  to  download  at 

www.cometsystem.com

.  It  supports  make  the  adjustment  of  the  device 

too.  This procedure  is  described  at  file  „

Calibration  manual.pdf

“  which  is  installed  commonly 

with the software.  Change  of  some  parameters  is  possible  to  do  without  user’s  software 
with Windows  hyperterminal  (change  of  communication  protocol,  its  parameters,  LCD  display 
setting).  It  is  described in file “

Description  of  communication  protocols  of  Txxxx series

” which is 

free to download at the same address. 
 

Transmitter  version  T

xxxx

L

  with  watertight  male  connector  instead  of  a  cable  gland 

(RS232) or glands (RS485) is designed for easy connection/disconnection of communication cable. 
Male Lumberg connector RSFM4 has protection IP67. 
 

Models  marked  T

xxxx

are  non-standard  versions  of  the  transmitters.  Description  is  not 

included in this manual. 
 

Please read instruction manual before the first device connection. 

 

 

Device setting from the manufacturer 

If special setting was not required in the order device is set from the manufacturer to the following 
parameters: 

 

communication protocol: 

Modbus RTU

 

device address: 

01H

 

communication speed: 

9600Bd, without parity, 2 stop bits

 

display:  

switched ON 

pressure unit:

 

hPa 

correction for altitude: 

0 hPa 

(absolute pressure) 

 
 
 
 

 
 

Summary of Contents for T2314

Page 1: ...cturer reserves the right to make technical changes to the device without previous notice Misprints reserved Manufacturer is not responsible for damages caused by using the device in conflict with thi...

Page 2: ...www cometsystem com It supports make the adjustment of the device too This procedure is described at file Calibration manual pdf which is installed commonly with the software Change of some parameter...

Page 3: ...r side The cable should be led in one line i e NOT to tree or star Termination resistor should be located at the end For short distance other topology is allowed Terminate the network by a termination...

Page 4: ...4 IE SNC T23 4 14 04 Dimensions T2314 Transmitter with RS232 Dimensions T2414 Transmitter with RS485 Dimensions T2314L T2414L Connection see Appendix B...

Page 5: ...IE SNC T23 4 14 04 5 Typical application wiring connection of terminals T2314 power from communication port T2314 power from external voltage source T2414 RS485...

Page 6: ...bit registers Read Holding Registers 04 0x04 Reading of 16 bit input gates Read Input Registers 16 0x10 Setting of more 16 bit registers Write Multiple Registers Jumper and button Jumper and button ar...

Page 7: ...g Note In case there is a need for reading of measured values from the device with higher resolution than one decimal measured values in device are stored also in Float format which is not directly co...

Page 8: ...ce with setup software TSensor to communication protocol HWg Poseidon and set correct device address This communication protocol supports read temperature at C relative humidity one of computed value...

Page 9: ...sure or error state of value If selected pressure unit is mBar or oz in2 or mmHg or inH2O or kPa there is shown only value number without corresponding pressure unit symbol 3 near by left display marg...

Page 10: ...allowed manipulation It is not allowed to operate the device under conditions other than specified in technical parameters Devices are not designed for locations with chemically aggressive environmen...

Page 11: ...connector marked RS232 directly to the PC connect power to connector marked RS485 Power voltage 6V DC from an external acdc adapter connect to pin 9 0V connect to pin 5 Also mutually connect pin 2 and...

Reviews: