It  is  recommended  to  use  shielded  twisted  copper  cable,  maximal  length  1200m.  The  cable 

must be located at indoor rooms. The cable should not be led in parallel along power cabling. Safety 
distance is up to 0.5 m, otherwise undesirable induction of interference signals can appear. Outside 
diameter of the cable for T2114 device must be from 3.5 to 8 mm (e.g. SYKFY), for device T2114L 
with respect to the female connector. Do NOT connect shielding at connector side. 

Electrical system (wiring) may do only worker with required qualification by rules in operation. 
 
 

Setting of pressure correction for altitude 

If there is necessary to read value of atmospheric pressure corrected with respect to altitude, 

it is possible to set corresponding pressure offset for this altitude. It is possible to do it with Users 
software. This value is then automatically added to measured pressure value. The resultant value of 
pressure with added correction for altitude must be between 300 hPa and 1350 hPa (from 4.351 PSI 
to 19.580 PSI). Outside this range device sends value of Error 2! 

 
 

Dimensions 

– T2114 

 

Connection: 

 
 
 
 
 
 
 
 
 

 

 

Dimensions 

– T2114L 

 

 
 
 
 
 

Connection:

 see Appendix A 

 
 
 
 

 

 
 

Summary of Contents for T2114

Page 1: ...turer reserves the right to make technical changes to the device without previous notice Misprints reserved Manufacturer is not responsible for damages caused by using the device in conflict with this manual To damages caused by using the device in conflict with this manual cannot be provide free repairs during the warranty period Read carefully instruction manual before the first device connectio...

Page 2: ...f special setting was not required in the order device is set from the manufacturer to the following parameters value at output I1 4 20 mA corresponds to atmospheric pressure from 800 to 1100 hPa1 correction for altitude 0 hPa absolute pressure display switched ON pressure unit hPa Modification of the setting is possible to do by means of the PC using procedure described at the end of this documen...

Page 3: ...side Electrical system wiring may do only worker with required qualification by rules in operation Setting of pressure correction for altitude If there is necessary to read value of atmospheric pressure corrected with respect to altitude it is possible to set corresponding pressure offset for this altitude It is possible to do it with Users software This value is then automatically added to measur...

Page 4: ...ted to the lowest pressure output value displayed at upper LCD line Here 4 mA corresponds to 800 hPa Press button again to get value for upper point similarly as at previous point Here 20 mA corresponds to 1100 hPa Notice during info mode no measurement and no output current generation proceed The device stays at info mode 15 s and then automatically goes back to measuring cycle Readings on LCD di...

Page 5: ...ering from the top side Electromagnetic compatibility complies EN 61326 1 Not allowed manipulations It is not allowed to operate the device under other than specified conditions in technical parameters Devices are not designed for locations with chemically aggressive environment Do not use the device in an explosive environment Storing conditions temperature 30 to 80 C humidity 0 to 100 RH without...

Page 6: ...ocedure to device s memory occurred or if damage of calibration data appeared At this state device does not measure and calculate values It is a serious error contact distributor of the device to fix Error 2 There is a reading Err2 on LCD display Atmospheric pressure measurement error occurred Output value is about 3 8 mA Measured pressure value with added correction for altitude is outside of ran...

Page 7: ...IE SNC T2114 03 7 Appendix A ...