IE-SNC-T2114-03
3
It is recommended to use shielded twisted copper cable, maximal length 1200m. The cable
must be located at indoor rooms. The cable should not be led in parallel along power cabling. Safety
distance is up to 0.5 m, otherwise undesirable induction of interference signals can appear. Outside
diameter of the cable for T2114 device must be from 3.5 to 8 mm (e.g. SYKFY), for device T2114L
with respect to the female connector. Do NOT connect shielding at connector side.
Electrical system (wiring) may do only worker with required qualification by rules in operation.
Setting of pressure correction for altitude
If there is necessary to read value of atmospheric pressure corrected with respect to altitude,
it is possible to set corresponding pressure offset for this altitude. It is possible to do it with Users
software. This value is then automatically added to measured pressure value. The resultant value of
pressure with added correction for altitude must be between 300 hPa and 1350 hPa (from 4.351 PSI
to 19.580 PSI). Outside this range device sends value of Error 2!
Dimensions
– T2114
Connection:
Dimensions
– T2114L
Connection:
see Appendix A