Sensor  operating  limits: 

the  electro-chemical  cell,  in  any  case,  should  not  be  exposed  to  temperatures  lower 

than -40° or higher than 60°C. The electrochemical cell should not  be exposed  to an over-range concentration  of 
30 % O

2

. If this condition occurs, we suggest replacing the sensor. 

Safety:

 Oxygen sensors contain a small quantity potassium acetate solution which is corrosive. Normally this mate-

rial would only leak out as a result of either mechanical damage or an incorrect instrument use beyond the limits of 
its operational functioning. These sensors also contain small amounts of lead, lead oxide, platinum, silver, carbon

 

and antimony. If the user comes into contact with the oxygen sensor contents the affected area should be washed 
with a copious supply of water. 

Cross  sensitivity  to  other  gases:

  The  electrochemical  cell  employed  has  a  good  resistance  towards  products 

such  as  sprays,  detergents,  ammonia,  glues  and  paints. 

However,  in  case  of  products  containing  substances  in 

great quantity, these could interfere with the Sensor and cause false alarms.

 We recommend ventilating the room 

when products like these are used. Volatile organic solvents in great quantity cause false alarms, and then the cell 
needs one or more hours in clean air to recover the normal sensitivity conditions. 

TEST and CALIBRATION 

PAY  ATTENTION

This  procedure  has  to  be  made  with  extreme  attention  and  by  authorized  and  trained 

people;  because  starting  this  procedure  it  will  start  both  Outputs  (relays)  causing  the  activation  of  con-
nected alarm devices. 

Operation Check, Calibration in Air and Fine mA ADJ Output:

 are different code protected functions. To access 

these functions is necessary to insert the relevant “

Code

" through the keys F1s and F2. To have the key pressure 

recognized, hold  pressing  it for around a second (

until the Green Led doesn't switch off for a moment

). Then the next key 

can be pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automatically erased. 
The 

Calibration Check

 is a function code free. 

Calibration Kit, Sample Gas Bottles

 (

for Calibration Check

) please, using a mixture: 

about 16%v/v Oxygen (O

2

) in Nitrogen

 

It is possible to use either the disposable cylinders with adjust valve or the high pressure ones with reduction gear. 

It is necessary to use 

the calibration kit 

model TC011

 

(

for non-corrosive gases

or TC014 (

Stainless steel

). 

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the equipment. After having put the system in safety and inserted the "

Code Test

", all Led are switched off and all 

the  relays  are  disarmed.  Then  they  will  switch  on  in  sequence,  the  Led,  from  the  yellow  up  to  the  3°  red.  To  the 
lighting  of  the  various  Led  it  corresponds  the  activation  of  the  relevant  relays  (

relay  "FAULT"  with  the  yellow  Led,  relay 

"ALARM1" with the 1° red Led, relay "ALARM2" with the 2° red Led, relay "ALARM3" with the 3° red Led

). At the end all the Led will 

remain lighted for around 5 seconds, then the central returns at the conditions of normal operation. It is advisable to 
perform this operation every 6-12 months according to the use. 

Note: this function is not working if the 1° and/or the 2° and/or the 3° red Led have already turned on 

 

“CALIBRATION IN AIR” (

Calibration Code

: F2, F2, F2, F1, F2, F1)

this function allows completely recalibrate the 

sensor at 20.9%v/v in air. This procedure when it is launched will automatically until the end. 

Warning

: to guarantee that no errors of elaboration happen, the rare possibility exists that during the Calibration 

the 

yellow LED switch off every 8 seconds, 

in this case interrupt the procedure, switch off and witch on the instru-

ment and repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the 
reparation. 

Important note:

 

During Calibration routine the mA output indicates 0mA

.

 

The Calibration can be done in clean air only (

environment without the presence of flammable or other polluting gas

). With the 

keys perform the “

Calibration

 Code

 ". Wait until the Yellow and Green Led switch on fix and the 1

st

 red Led starts to 

flash (the 3

th

 Led Red

 

switch on to indicate the calibration working). Wait while the 1

st

 red Led switched off. At this 

point we can have two possibilities: 

Yellow and Green Led illuminates:

 the calibration routine has correctly been performed. The detector automatically 

switches off in 8 seconds, and restart into normal operating mode. (

See chapter OPERATIONAL DESCRIPTION “Preheating”

). 

Yellow LED illuminates:

 the calibration routine has failed. The detector automatically switches off and restarts in 

8  seconds  and  repeats  the  procedure  without  inserting  again  the  sequence.  If  condition  still  persists  after  the  re-
placement of the cartridge it will be necessary to send the detector back to the manufacturer for reparation. 

Important

:  If  the  environment  is  much  polluted  and  there  is  doubt  that  the  concentration  of  oxygen  is  20.9%v/v, 

you can use an air cylinder with 

20.9%v/v of O

2

 or use a gas balloon filled with fresh air. In this case, before exe-

cuting the  "

Calibrating Code

"  inserts the TC011 over the  sensor holder, adjust the flow of air, so that the flow  meter 

indicates approximately 0.3 l / min (

see Fig.3

). 

“CALIBRATION  CHECK”  (

no  Code  required

)

:

 

although  the  sensor  operation  is  guaranteed  if  it  is  regularly  per-

formed successfully "Calibrating in Air", it is possible to simulate the response of the detector using sample gas cyl-
inder  with  a  mixture  16%v/v  Oxygen  in  nitrogen.  Insert  the  TC011  over  the  sensor  holder;  adjust  the  sample  gas 
cylinder valve as the flow meter indicates around 0.3 l/mins (

see Fig.3

), wait for 1÷2 minutes or until the instrument 

activates Alarms as shown in chapter “

Operational Description

”. 

If the result is different, is necessary to recalibrate the 

sensor (

see “Calibration in Air”

).

 Then, close the gas cylinder

 

and remove TC011, 

wait 2÷3 minutes, until the instru-

ment restores the normal working conditions

NOTE

:  further  to  the  above,  if  the  mA  output  is  also  used,  verify  with  a  voltmeter  connected  to  the  Test-Point 

TESTmA

” (

see Fig.2

), the value reach a 

174mV

 value in normal clean air condition (20.9%v/v). 

Summary of Contents for 47RTS282EO

Page 1: ...a Response Time Temps de r ponse T90 20 secondi seconds secondes Ripetibilit Repeatability R p titivit 1 del segnale signal Precisione Accuracy Precision 2 del segnale in aria signal in air du signal...

Page 2: ...FAULT Dopo circa 60 secondi il Led giallo si spegne e si accende il Led verde ON che indica il normale funzionamento Dopo questo tempo il sensore in grado di rilevare il gas ma raggiunge le condizion...

Page 3: ...ni di O2 molto elevate oltre il 30 In questi casi il sensore deve essere sostituito Attenzione i sensori elettrochimici per Ossigeno contengono una piccola quantit di acetato di potassio che corrosivo...

Page 4: ...on i puntali del milliamperometro direttamente sui morsetti S e verificare che ci sia 17 4mA circa Applicando il gas al 15 O2 controllare che si raggiunga un valore di circa 136mV ovvero che l uscita...

Page 5: ...activate and 0mA output signal this happens when the Cartridge Sensor is not working First try to perform the procedure of Calibration in Air as described in the section Test and Calibration Calibrati...

Page 6: ...red Led relay ALARM3 with the 3 red Led At the end all the Led will remain lighted for around 5 seconds then the central returns at the conditions of normal operation It is advisable to perform this...

Page 7: ...er les op rations de v ri fication et calibrage en utilisant par codes les touches F1 et F2 CARACT RISTIQUES DU GAZ L Oxyg ne O2 est un gaz incolore inodore l g rement plus lourd que l air sa densit r...

Page 8: ...t indi qu e dans le Tableau 1 en fonction de la section du c ble utilis Dans le cas d alimentation de plusieurs d tecteurs en parall le il est n cessaire de calculer la chute de tension sur les portio...

Page 9: ...squ la fin Durant la proc dure une rare possibilit existe que la LED Jaune clignote chaque 8 secondes dans ce cas inter rompre la proc dure puis ex cutez le contr le en coupant et en r tablissant l al...

Page 10: ...lla to ogni rivelatore se alimentato a 12Vcc The maximum distance to install each detec tor from the 12Vdc power Supply Distance max d installation d un d tecteur sous 12Vcc Massima distanza cui pu es...

Page 11: ...a di collegamento Wiring diagram Sch ma TESTmV TP2 TP1 Scheda Uscite Rel Relays Output Board Sortie sur Realis S 12 24V mA 12 24Vdc Alimentazione Power Supply Alimentation Fig 3 Tester di calibrazione...

Reviews: