Error Messaging
12
Message
Parameters
Error type
FAiL
all
incorrect memory
br"
-
service required
ER01 GrAd
out of range
ER02 \t
out of range
ER03 \b
out of range
ER04 %[L
out of range
ER05 %[H
out of range
ER06
%[L
,
%[H %[L
>
%[H
ER11
,
ER21 ym1
,
ym2
out of range
ER12
,
ER22 y\1
,
y\2
out of range
ER13
,
ER23 HLe1
,
HLe2
out of range
ER14
,
ER24 [e1
,
[e2
out of range
ER15
,
ER25 me1
,
me2
out of range
ER16
,
ER26 %[1
,
%[2
outside
%[L
...
%[h
ER17
,
ER27 %[
x
-
me
x
under input range
ER18
,
ER28 %[
x
+
[e
x
over input range
ER29 Addr
out of range
♦
In some cases, TC67U
finds discrepancies
in parameter values
that must be resolved
before operating
at Basic level.
♦
The device indicates
such kind of problems
by displaying
error messages
as given on the left.
♦
If
FAiL
appears
on the display,
try debugging
by turning
the power off/on.
♦
If the problem persists,
press and hold
or
send command
error
0
via the communication
interface to restore
the default (factory)
settings.