3-PHASE LINE IMPEDANCE STABILIZATION NETWORKS (LISNs)
SECTI ON 6 -
LISN MEASUREMENTS
1912 1 E l T o ro R d
●
Si l vera do, Ca lifo rni a 9 26 76
●
(9 49) 459 -96 00
●
co m-p o wer. com
REV031418
6.1.3
Non-LISN Insertion Loss Calibration
Insertion Loss values for coaxial cables and most measurement system
components having a single coaxial input and output, such as
attenuators, filters, dc blocks, etc., can be easily determined through a
simple calibration process.
The test setup for normalization and insertion loss measurements are
illustrated in Figure 12. The insertion loss factor is the difference (in dB)
between the normalization measurements and the insertion loss
measurements.
FIGURE 12 -
Setup Diagrams for Insertion Loss Calibration
Insertion Loss Calibration
NORMALIZATION MEASUREMENTS
Insertion Loss Calibration
INSERTION LOSS MEASUREMENTS
INPUT
RF
Network Analyzer
with
S-Parameter Test Set
- or -
Synthesized RF Signal Generator
and Spectrum Analyzer/Receiver
OUTPUT
RF
INPUT
RF
Network Analyzer
with
S-Parameter Test Set
- or -
Synthesized RF Signal Generator
and Spectrum Analyzer/Receiver
OUTPUT
RF
system component
to be calibrated
10 dB
10 dB
X
10 dB
10 dB