INSTRUCTION MANUAL
SECT IO N 4 -
PRODUCT INSTALLATION
1912 1 E l T o ro R d
●
Si l vera do, Ca lifo rni a 9 26 76
●
(9 49) 459 -96 00
●
Rev050321
The ADA-515-CLEM 150 ohm to 50 ohm
Adapters
and
the
CLEMA-6146
Calibration Adapters allow for adjustment
of the height of the 4 mm copper rod. For
each adapter set, the height adjustment
is made by loosening the four screws
shown circled in
red
on Figure 10,
adjusting the height up or down, then re-
tightening the screws.
For
coupling factor
and
test level calibration
measurements
using the ADA-515-CLEM Adapters, the height should be
adjusted so that the 4 mm rod is located in the
bottom position
of the clamp opening, as shown in Figure 11.
For
impedance
and
decoupling factor
measurements using
the CLEMA-6146 Adapters, the height should be adjusted so
that the 4 mm rod is located in the
center position
of the
clamp opening, as shown in Figure 11.
It is critical that the EM Clamp be installed in a manner which ensures that EACH of
following conditions are satisfied:
•
The metallic mounting bracket must be connected to earth ground.
•
The input signal power must not exceed the ratings/time durations specified in
the product specifications table in section 3.5.
Failure to comply with any of these points may damage the equipment and/or
pose an electrical hazard.