Calibration fixtures provide a means by which current probes can be calibrated to
determine the insertion loss and transfer impedance factors for the probe. Fixtures are also
used when establishing or calibrating, drive levels for conducted susceptibility tests. These
applications are discussed in more detail in the following sections.
The FCLCE-1000 Fixture is specifically designed for use with the CLCE-400.
The intent of the fixture is to maintain the coaxial structure of the transmission line, while
allowing the probe to be installed around the center conductor of the coaxial line.
Illustrated in Figure 8 is the procedure to be followed for installing the CLCE-400
Current Probe into the FCLCE-1000 Calibration Fixture.