2. Hollow Wrapped Cycles
The Statim 2000S and 5000S
feature two WRAPPED 134˚C type S
sterilization cycles.
To select one of these S Cycles:
Press the WRAPPED button to
scroll through the available cycles.
Once the desired cycle has been selected, press the START button.
The unit remembers the last cycle run and then displays this first when the unit
is powered on.
A Process Challenge Device (PCD) is available for validation of HOLLOW WRAPPED (S)
134˚C/3.5 min cycle.
3. Rubber and Plastics Cycle
The Statim 2000S and 5000S feature
two
121˚C type S sterilization cycles.
To select one of these S Cycles:
Press the RUBBER / PLASTIC button
to scroll through the available cycles.
4. Instructions For Use Con’t.
A
B
C
D
E F
D Sterilizing
134˚C / 304 kPa -
138˚C / 341 kPa
E
Venting
F
Air Drying
3.5
min /
18 min
A 100˚C / 101 kPa
B 6 purges (Conditioning)
peak at 133˚C / 295 kPa
vent to 115˚C / 169 kPa
C
Pressurizing
HOLLOW WRAPPED (S) 134˚C / 3.5 min
HOLLOW WRAPPED (S) 134˚C / 18 min
A
B
C
D
E
F
D Sterilizing
121˚C / 205 kPa -
124˚C / 225 kPa
E
Venting
F
Air Drying
15
min /
30
min
A 100˚C / 101 kPa
B 3 purges (Conditioning)
peak at 120˚C / 199 kPa
vent to 110˚C / 143 kPa
C
Pressurizing
RUBBER / PLASTIC (S) 121˚C / 15 min
RUBBER / PLASTIC (S) 121˚C / 30 min