SC Scanners Scanner Service Manual

 

Page 4 of 7 

3.  Continue to gently pull the 

whole SingleSensor Glass 
Sub-Assy away from the 

Error! Reference source 
not found.

 

 

4.  Remove the SingleSensor 

Glass Sub-Assy 

 

 

5.  When refitting the 

SingleSensor Glass Sub-
Assy, make sure the right 
end is against the End Plate. 

 

 

6.  Position the SingleSensor 

Glass Sub-Assy over the 

Error! Reference source 
not found.

 and make sure it 

is central all the way along. 

 
 

NOTE: Before fitting the SingleSensor Glass 
Sub-Assy, clean both sides with a lint free 
cloth to remove any dust, dirt or fingerprints. 

 

 

Summary of Contents for SmartLF SC25c

Page 1: ...ion Manual Date 12 02 14 This is a confidential controlled document and is the property of Colortrac Ltd It is not to be copied or reproduced in part or whole C Co ol lo or rt tr ra ac c S SC C S Sc c...

Page 2: ...or Misprints errors or quality observations should be reported to Technical Support Colortrac Ltd Colortrac Ltd shall not be liable for any damages errors issues or events that result from the use of...

Page 3: ...maintenance platform 1 1 2 Tools Medium Flat Screwdriver Soft Cloth 1 1 3 Associated Part Numbers 10H014 SingleSensor Glass Sub Assy 1 1 4 Procedure 1 To Remove Open the Error Reference source not fou...

Page 4: ...sor Glass Sub Assy 5 When refitting the SingleSensor Glass Sub Assy make sure the right end is against the End Plate 6 Position the SingleSensor Glass Sub Assy over the Error Reference source not foun...

Page 5: ...ensor Glass Sub Assy until all the latches are locked in place 8 Close the Error Reference source not found 9 Reconnect the Power and USB cables to the Scanner and power up the scanner and Utilities P...

Page 6: ...ce source not found and clean the SingleSensor Glass Sub Assy with a soft cloth Close the Lid Assembley 2 Insert the Calibration Target so that the black part section is inserted first face up NOTE It...

Page 7: ...er Service Manual Page 7 of 7 5 Click OK to except the message 6 After approx 4 minutes the Calibration or Normalisation is complete During calibration the target will move through the scanner in stag...

Reviews: