INSTALLATION

5.1.   Electrical safety and Installation

5.1.1  This equipment is designed for safe operation under the following conditions:-

• Indoor use.

• Altitude up to 2000 meters.

• Temperatures between 5°C and 40°C.

• Maximum relative humidity 80% for temperatures up to 31°C decreasing
   linearly to 50% relative humidity at 40°C.

• Mains supply voltage fluctuations up to ± 10% of the nominal voltage.

• Transient overvoltages typically present on the mains supply.  

• Applicable rated pollution degree 2.

5.2.    

   This equipment must be earthed / grounded to a fixed earth / grounded mains

 

socket outlet. The mains supply is to be earthed / grounded in accordance with

 

current legislation.

5.3.   Ensure only the correct rated mains input fuses are fitted. (Where applicable ensure 

the correct Mains cord and moulded IEC plug and lead set fuse if fitted). See Technical 
Specification Section 8 of this Instruction manual.

5.4.   Check the input voltage on the product data label of each Digi-Mantle item. Ensure
 

the rating conforms to your local supply.

5.5.   This product should be connected to a mains supply source which incorporates an
 

RCD or GFCI device that has a tripping current of 30mA or less. The RCD or GFCI

 

residual Current Device cuts off power to the equipment immediately it detects a

 

current leakage fault. For example, cutting off the power when there is an accidental

 

liquid spillage in a mantle protected with an earth (ground) screen. 

5.6.   The unit is supplied with a moulded mains cord and plug set wired as follows:-

 

Green /Yellow or Green = Earth / Ground

 

Blue or White = Neutral

 

Brown or Black = Live / line hot.

 

When using with a new mantle:

 Observation:

 the surface of the heating element of a mantle cartridge will upon receipt 

look slightly discoloured. This discolouration is normal and occurs at the factory during 
testing when the mantle is first heated up.

Page 8 of 24                                                                                                                                  M6493 Issue 10.3 

 
5.  INSTALLATION. 
 

5.1.  Electrical safety and Installation. 

 

5.2.  This equipment is designed for safe operation under the following conditions:- 

 

 

Indoor use. 

 

 

Altitude up to 2000 meters. 

 

 

Temperatures between -10

°

C and +50

°

C. 

 

 

Maximum relative humidity 80% for temperatures up to 31

°

C decreasing 

linearly to 50% relative humidity at 40

°

C. 

 

 

Mains supply voltage fluctuations up to 

±

 10% of the nominal voltage. 

 

 

Transient overvoltages typically present on the mains supply. (Overvoltage 
category II). 

 

 

Applicable rated pollution degree 2. 

 

5.3. 

 This equipment must be earthed / grounded to a fixed earth / grounded mains 

socket outlet. The mains supply is to be earthed / grounded in accordance with 
current legislation.  

 
5.4.  Ensure only the correct rated mains input fuses are fitted. (Where applicable ensure 

the correct mains cable fuse if fitted). See Technical Specification, Section 8 of this 
Instruction book.  

 

5.5.  Check the voltage on the product data label of this unit. Ensure the rating conforms 

to your local supply. Only connect to resistive loads = 1500Watts @ 230V~AC or 
750Watts @ 115V~AC. Refer to Technical Specification, Section 8 of this Instruction 
book.  

 

5.6.  This product must be connected to a mains supply source which incorporates an 

RCD or GFCI device. 

 

5.7.  The unit is supplied with a moulded mains cord and plug set wired as follows:- 

 

 Green / 

Yellow 

 
or 

 

Green 

 

 
Earth / Ground 

 

 

 

 

 

Blue 

or 

White 

Neutral 

 

 

 

 

 

Brown 

or 

Black 

Live / line hot. 

 

 

 

 

 

5.8.  If the controller is to be used on a retort rod, make sure it isn’t positioned over any 

heating apparatus. 

 

5.9. 

If using the wall mounting bracket provided, carefully mark and drill two 

suitable holes using the bracket as a template. Ensure there are now cables etc 
buried in the wall. Plug the holes with raw plugs and fasten the bracket to the wall in 
the correct orientation using the screws provided. Place the controller over the 
bracket. 

Page 8 of 24                                                                                                                                  M6493 Issue 10.3 

 
5.  INSTALLATION. 
 

5.1.  Electrical safety and Installation. 

 

5.2.  This equipment is designed for safe operation under the following conditions:- 

 

 

Indoor use. 

 

 

Altitude up to 2000 meters. 

 

 

Temperatures between -10

°

C and +50

°

C. 

 

 

Maximum relative humidity 80% for temperatures up to 31

°

C decreasing 

linearly to 50% relative humidity at 40

°

C. 

 

 

Mains supply voltage fluctuations up to 

±

 10% of the nominal voltage. 

 

 

Transient overvoltages typically present on the mains supply. (Overvoltage 
category II). 

 

 

Applicable rated pollution degree 2. 

 

5.3. 

 This equipment must be earthed / grounded to a fixed earth / grounded mains 

socket outlet. The mains supply is to be earthed / grounded in accordance with 
current legislation.  

 
5.4.  Ensure only the correct rated mains input fuses are fitted. (Where applicable ensure 

the correct mains cable fuse if fitted). See Technical Specification, Section 8 of this 
Instruction book.  

 

5.5.  Check the voltage on the product data label of this unit. Ensure the rating conforms 

to your local supply. Only connect to resistive loads = 1500Watts @ 230V~AC or 
750Watts @ 115V~AC. Refer to Technical Specification, Section 8 of this Instruction 
book.  

 

5.6.  This product must be connected to a mains supply source which incorporates an 

RCD or GFCI device. 

 

5.7.  The unit is supplied with a moulded mains cord and plug set wired as follows:- 

 

 Green / 

Yellow 

 
or 

 

Green 

 

 
Earth / Ground 

 

 

 

 

 

Blue 

or 

White 

Neutral 

 

 

 

 

 

Brown 

or 

Black 

Live / line hot. 

 

 

 

 

 

5.8.  If the controller is to be used on a retort rod, make sure it isn’t positioned over any 

heating apparatus. 

 

5.9. 

If using the wall mounting bracket provided, carefully mark and drill two 

suitable holes using the bracket as a template. Ensure there are now cables etc 
buried in the wall. Plug the holes with raw plugs and fasten the bracket to the wall in 
the correct orientation using the screws provided. Place the controller over the 
bracket. 

Summary of Contents for SHM-250 Series

Page 1: ...Cole Parmer SHM 250 Series Digi Mantle Instruction Manual M7754 Version 14 6...

Page 2: ...ction manual is the property of Antylia Scientific Limited This instruction manual is supplied by Antylia Scientific Limited on the express understanding that it is to be used solely for the purpose f...

Page 3: ...the heat setting as a percentage of the total power When the temperature probe is fitted the heating is controlled via the temperature feed back circuit and the actual temperature is displayed All ne...

Page 4: ...estic waste Caution Hot surface Refer to Operator Instruction Manual Stirring Facility On Off Heating Facility On Off Bio Chemical Hazard Caution required Will require decontamination 4 YMBOLS AND USI...

Page 5: ...ractice Always follow good laboratory practice when using this equipment Give due recognition to your company s safety and legislative health safety procedures and all associated legislation applicabl...

Page 6: ...perating Practice Always follow good laboratory practice when using this equipment Give due recognition to your company s safety and legislative health safety procedures and all associated legislation...

Page 7: ...rope 1x Mains lead and moulded IEC plug and lead set USA 1x Mantle 0250ml OR 0500ml OR 1000ml 1x Digi controller with inbuilt stir 4 UNPACK AND CONTENTS 4 1 OMCA0250 OMCA0250X1 OMCA0500 OMCA0500X1 OMC...

Page 8: ...I Applicable rated pollution degree 2 5 3 This equipment must be earthed grounded to a fixed earth grounded mains socket outlet The mains supply is to be earthed grounded in accordance with current le...

Page 9: ...issues within the design and manufacturing process without compromising end product performance and value 6 2 Packaging materials have been selected such that they may be sorted for recycling 6 3 At...

Page 10: ...tion key 3 Heater activation key 4 LCD display 5 Menu selection keys Figure 7 2 6 On off power switch 7 PRODUCT OPERATION 7 1 Overview Your Digi Mantle has been designed for easy operation The illustr...

Page 11: ...Data communication 17 Read instruction book Figure 8 Typical Front Screen Display Figure 7 4 Figure 7 3 7 External probe socket 8 Date plate label 9 Dry block control plug 10 Dry block power supply so...

Page 12: ...item 6 to the on position 7 4 2 The LCD display will go through a warm and self check routine that will take a few seconds during which the Digi Mantle Software revision will be displayed for 5 secon...

Page 13: ...ed selection screen See figure 7 9 Figure 7 9 Press the key to make the 100 s value selection Press the v key if you wish to reduce the selected stir speed value Press the key to go into the 10 s sele...

Page 14: ...turn to the main display menu To start the heating process press the heating key marked with the symbol To stop the heating process press the heating key once more Note To show heating is in operation...

Page 15: ...n Heating your sample is now controlled via the external probe From the main display menu screen press the key to go into the Temperature setting screen See figure 7 13 Figure 7 13 Note the Stir tempe...

Page 16: ...ssue 14 3 7 7 Timer Operation Your Digi Mantle has the facility to set a countdown timer which will switch off the heaters after a per set time period The timer can be set for up to 99hours 59 minutes...

Page 17: ...the main display screen reverts to t A one second beep is sounded when the heaters are turned off in timer mode Note At any time during the use of the timer facility the heaters may be turned off man...

Page 18: ...figure 7 29 Figure 7 29 The unit will beep rapidly for 10 seconds The heaters will be turned off If the unit was stirring a sample it will continue to do so The Warning will remain on the LCD until th...

Page 19: ...in 60 seconds the heaters will be turned off The unit will beep rapidly for 10 seconds If the unit was stirring a sample it will continue to do so The Warning will remain on the LCD until the key is...

Page 20: ...1000 230V 300W Stir speed Min 100 RPM Max 2000 RPM Fuse rating 115V 7A 32mm x 6 3mm Quickblow 230V 5A 32mm x 6 3mm Quickblow Product configuration details SHM 250D 250 115 115V Controller with stirrer...

Page 21: ...oller and mantle Weight 3 1kg 21 8 3 Dimensions Weight unpacked Controller Dimension Weight unpacked Controller and Mantle D 330mm W 235mm H190mm Weight 3 1Kg 8 3 Dimensions Weight unpacked Controller...

Page 22: ...le rating and value See section 8 1 for correct type rating and value and refit into fuse housing See Figure 9 1 Figure 9 1 The equipment is not spark flame or explosion proof and has not been designe...

Page 23: ...n by pressing button 9 5 Spillage and Decontamination In the event of spillage switch off and unplug this product from the mains electrical supply Wipe off all excess liquid from the unit and surround...

Page 24: ...X1 1000ml size mantle 300W 115V HH179 S Mains cord and moulded IEC plug and lead set cable UK 10A BS1362 HH180 S Mains cord and moulded IEC plug and lead set cable Europe CRM6288 Mains cord and moulde...

Page 25: ...r Services at the following address Antylia Scientific Ltd Beacon Road Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 1785 812121 Service cpservice antylia com General enquiries cpinfo antylia c...

Page 26: ...26...

Page 27: ...o we cannot guarantee that interference will not occur in practise Where there is a possibility that injury damage or loss might occur if equipment malfunctions due to radio frequency interference or...

Page 28: ...a T 9122 61394444 E info coleparmer in W coleparmer in China T 1 847 549 7600 E sales antylia com W coleparmer com USA T 1 847 549 7600 E sales antylia com W coleparmer com Canada T 514 355 6100 E inf...

Reviews: