6
SETUP AND DRIVE OPERATION
1. Unpack the drive: save packaging material until proper product operation
has been verified.
2. Select and mount Pump Head and load tubing (see Pump Head manual).
3. For 77200-00 and 77200-10 plug in External Power supply and connect to
the back of the drive. (See Figure #3.)
4. For 77200-12 supply 11.0 to 15V DC to terminal strip positions 1 and 2.
(See Figure #3.)
5. The drive system can be set to operate in either a clockwise or a counter-
clockwise direction. The same control used to select direction also turns
power on or off. To operate in a clockwise direction place direction control
switch in the “up” position. The POWER indicator should light and the fluid
will flow from right to left. To operate in a counterclockwise direction, place
direction switch in the “down” position. The POWER indicator should light
and the fluid will flow from left to right. To turn pump off, place switch in the
“center” position.
6. The pump speed is controlled by the variable speed control. Turning the
control clockwise increases the speed. Tubing life is decreased with in-
creased operating speed. Speed can be controlled up to a 6:1 range.
MAINTENANCE
Cleaning: Keep the drive enclosure clean with mild detergents. Never im-
merse nor use excessive fluid.
Motor/ Brush Replacement: The motor used in this inexpensive drive is not
serviceable or repairable and has brushes which cannot be replaced. The
motor is rated for a continuous duty life of 3000 operating hours minimum.
Operation at other than continuous duty cycles will result in longer motor life.
If motor failure for whatever reason occurs, the motor must be replaced.