background image

BLOCK HEATERS

BH-200-2, BH-200D-2, BH-200D-2-HT, 
BH-200D-3, BH-200D-3-HT BH-200DC-2 and 
BH-200D-2-HTC.

For your own safety and that of others 
please read and understand the safety 
advice given in section 9 of this manual 
before using the equipment.

If the equipment is not used in the manner 
described in this manual the protection provided 
might be impaired.

1. General  description

Cole-Parmer block heaters are designed to 
accurately control the temperature of test tubes 
and other small containers using heated 
aluminium blocks drilled with holes of the 
appropriate size for the containers in use.  A 
large variety of blocks with different holes for 
different sizes and type of container are available 
– see section 7.
The blocks are fitted into the recess in the top of 
the instrument (see figure 1) and heated from 
below. All blocks are fully interchangeable and 
blocks for different sizes of container may be 
used in a block heater at the same time.
The large thermal mass of the aluminium blocks 
allows very accurate and stable temperature 
control.
Perspex safety covers are available as accessories 
for protection in the event of tubes “spitting”.  
These also help improve temperature uniformity 
within the blocks by excluding draughts.

2. Preparation for use – 

  All models

1.   Follow the instructions for Electrical 

Installation given in section 8 of this manual 
and connect the instrument to the electricity 
supply – 

DO NOT SWITCH ON.

2.  Choose aluminium blocks with holes suitable 

for the containers to be heated.

3.  Carefully place the appropriate number of 

aluminium blocks into the recess in the top of 
the instrument – see figure 1.

1

N.B.  The plate at the bottom of the recess 
and the base of the aluminium blocks must 
be kept scrupulously clean to allow good 
thermal contact.  If foreign matter, especially 
particulate matter, is present the 
performance of the instrument will be 
impaired.

4.  Place the containers to be heated into the 

holes in the aluminium blocks.

5.  Fit safety cover if required.  This may simply 

rest on the top of the block heater or be 
hinged by using the screws provided.  These 
should be located in the holes at the top rear 
of the block heater (see figure1).

6.  The instrument may now be switched on and 

the desired temperature set.

  

3. Temperature  setting

3.1

 

Digital models BH-200D-2, BH-200D-2-HT, 
BH-200D-3, BH-200D-3-HT, BH-200DC-2,

 

and BH-200D-2-HTC.

1.   Switch on using the switch located at the rear 

of the instrument – see figure 2.

2.   The “power” light will illuminate as will the 

digital display - see figure 3.

3.   The digital display will show the actual 

temperature of the aluminium blocks.  If the 
instrument’s current set temperature is above 
this value it will begin to heat and the “heat” 
light on the control panel will illuminate. The 
current set temperature may be viewed at 
any time by pressing and holding the “set” 
button – see figure 3.

4.   To adjust the set temperature first press and 

hold the “set” button to show the current 
setting.  

Figure 3 - digital controls

Summary of Contents for BH-200 Series

Page 1: ...Cole Parmer BH 200 Series Block Heaters Instruction Manual SBH061 ES Version 1 6...

Page 2: ...e Parmer equipment To get the best performance from the equipment please read these instructions carefully before use Figure 1 Front view Figure 2 Rear view Block extraction tool Recess for blocks On...

Page 3: ...to the electricity supply DO NOT SWITCH ON 2 Choose aluminium blocks with holes suitable for the containers to be heated 3 Carefully place the appropriate number of aluminium blocks into the recess in...

Page 4: ...peration of the controls for each block is exactly as described above If using this model with only one block in place the side with no block should have its set temperature adjusted to 30 C 3 3 Analo...

Page 5: ...he unit will begin to heat 6 After approximately 30 minutes the unit will have stabilized at approximately 40 C and the display will change from showing temperature to read EEEE 7 Accurately measure t...

Page 6: ...eter if desired All blocks have dimensions w x d x h of 75 x 95 x 50mm and can be used in any combination The following accessories are available from Antylia Scientific or its agents Cat No Descripti...

Page 7: ...perature range 5 C to 40 C Altitude to 2000m Relative humidity not exceeding 80 Mains supply fluctuation not exceeding 10 Over voltage category II IEC60364 4 443 Pollution degree 2 Use with a minimum...

Page 8: ...r en obtenir les meilleures performances veuillez lire attentivement ces instructions avant de l utiliser Figure 1 Vue de face Figure 2 Vue arri re Outil d extraction des blocs R ceptacle des blocs In...

Page 9: ...section 8 du pr sent manuel et raccorder l instrument l alimentation lectrique NE PAS METTRE SOUS TENSION 2 Choisir des blocs d aluminium ayant des trous adapt s pour les r cipients chauffer 3 Placer...

Page 10: ...r L appareil devra ensuite tre d branch de l alimentation lectrique Noter qu la prochaine utilisation l instrument aura conserv sa temp rature de consigne pr c dente 3 2 Mod le num rique BH 200DC 2 BH...

Page 11: ...cs ne peut qu avoir une pr cision gale celle du thermom tre utilis pour l talonner Suivre la proc dure d finie ci dessous 1 Commencer avec des articles froids 2 Les blocs devront tre fermement pouss s...

Page 12: ...our prot ger les blocs du courant d air 5 Lors de l talonnage la sonde de temp rature utilis e devra s adapter parfaitement dans le trou du bloc 6 Retrait des blocs d aluminium N B Ne pas essayer de r...

Page 13: ...nt livr s avec deux cordons secteur quip s de fiches IEC permettant le branchement l appareil Un cordon poss de une fiche 3 broches britannique et l autre est quip d une fiche 2 broches Schuko permett...

Page 14: ...CHAUDS Ne jamais d placer ni transporter l appareil sans l avoir d branch de l alimentation lectrique et laiss refroidir Les blocs ne devront JAMAIS tre retir s de l instrument avant que celui ci ait...

Page 15: ...di ottenere le prestazioni ottimali si invita l utente a leggere attentamente le presenti istruzioni prima dell uso Figura 1 Vista anteriore Figura 2 Vista posteriore Estrattore portaprovette Vano per...

Page 16: ...a leggere e osservare le norme di sicurezza riportate nella sezione 9 del presente manuale prima di utilizzare l apparecchio L inosservanza delle istruzioni riportate nel presente manuale durante l u...

Page 17: ...minio a questa temperatura fino a quando non si modifica il valore impostato o non si spegne l apparecchio 11 possibile estrarre o inserire contenitori dal vano in qualsiasi momento ad apparecchio in...

Page 18: ...tro dell apparecchio stesso poi staccare la spina dalla presa elettrica 4 Taratura Solo per i modelli digitali Questi apparecchi vengono tarati durante la fabbricazione utilizzando dei blocchi destina...

Page 19: ...lmente 5 Fattori che influiscono sulla taratura e sulla precisione Al fine di ottenere la massima precisione dall apparecchio e di assicurare la precisione della taratura osservare i provvedimenti ind...

Page 20: ...figura 2 Avvertenza I fusibili sono montati sulla linea di fase e neutra Gli apparecchi vengono forniti con due cavi di alimentazione elettrica muniti di spine IEC Un cavo munito di una spina tripolar...

Page 21: ...o collegato alla rete elettrica Non accendere o riscaldare l apparecchio se non sono stati inseriti i blocchi di alluminio previsti nel vano previsto nella parte superiore dell apparecchio I blocchi d...

Page 22: ...loques Cavidad para introducir los bloques Interruptor de encendido apagado Recept culo para los fusibles Panel de control Conector IEC Le agradecemos que haya adquirido este equipo Cole Parmer Para c...

Page 23: ...o existen unas cubiertas de seguridad fabricadas en Perspex metilmetacrilato que pueden utilizarse para proteger el equipo de posibles derrames de los tubos Adem s estas cubiertas ayudan a mejorar la...

Page 24: ...os bloques de aluminio hasta alcanzar la temperatura indicada 9 La luz heat del panel de control permanecer encendida durante todo el proceso de calentamiento y parpadear cuando se aproxime a la tempe...

Page 25: ...a se puede dejar la rueda en su posici n y utilizar el interruptor de la parte trasera para encender y apagar el equipo Tenga en cuenta que si apaga el equipo usando el interruptor la luz de advertenc...

Page 26: ...ue tanto la temperatura mostrada en pantalla como la temperatura obtenida con el term metro est n entre 124 C y 126 C o entre 194 C y 196 C seg n del modelo 20 Pulse el bot n set para confirmar la nue...

Page 27: ...ra cubetas de cristal y desechables SHT1 22 Tubos de 1 5 ml 20 10 7 x 14 SHT1 25 25 mm 06 25 5 x 47 SHT1 28 28 mm 06 28 0 x 47 SHT1 30 Tubos de 0 5 ml 30 7 8 9 de inclinaci n SHT1 30 1 30 mm 04 30 1 x...

Page 28: ...ctar el equipo lea detenidamente estas instrucciones y aseg rese de que el suministro el ctrico corresponde con el que se muestra en la siguiente lista de clasificaci n El consumo de energ a es el sig...

Page 29: ...l quido en los bloques deber desconectar el equipo inmediatamente y dejarlo enfriar Antes de volver a usar el equipo deber limpiar y secar los bloques y la cavidad superior del equipo Aseg rese de que...

Page 30: ...er Bl cke Aussparung f r Bl cke Ein Aus Schalter Sicherungskasten Steuerkonsole IEC Anschluss Vielen Dank f r Ihre Entscheidung zum Kauf dieses Ger tes von Cole Parmer Um die optimale Leistung des Ger...

Page 31: ...ung Perspex Sicherheitsabdeckungen sind als Zubeh r erh ltlich und dienen zu Ihrem Schutz im Fall dass Fl ssigkeit aus den R hrchen verspritzt wird Diese Abdeckungen vermeiden einen Luftzug im Ger t u...

Page 32: ...e Anzeige auf die Darstellung der aktuellen Temperatur umzuschalten 8 Das Instrument beginnt jetzt damit die Aluminiumbl cke auf diese Temperatur aufzuheizen 9 Die Lampe mit der Aufschrift Heat auf de...

Page 33: ...Glasgef e in die vorgeheizten Bl cke stellen m ssen Sie Vorsicht walten lassen da die Beh lter aufgrund des Hitzeschocks zerbrechen k nnten 8 Falls die Temperatur exakt eingestellt werden muss kann si...

Page 34: ...zu best tigen Die bernsteinfarbene Lampe Heat h rt auf zu blinken und leuchtet dauerhaft auf 11 Die Selbstkalibrierung wird jetzt fortgesetzt und das Ger t heizt bis auf ca 5 C unter den Maximalwert a...

Page 35: ...kzeug f r die Entnahme der Bl cke versehen Dieses Spezialwerkzeug wird in das Thermometerloch oben auf dem Block geschraubt Anschlie end kann der Block aus dem Instrument gehoben werden F r die Verwen...

Page 36: ...Instrument ausger stet Ein Kabel hat einen dreipoligen britischen Stecker und das andere hat einen zweipoligen Schuko Stecker f r den Netzanschluss W hlen Sie das entsprechende Kabel f r Ihre Stromver...

Page 37: ...n der Stromversorgung getrennt worden ist und sich abgek hlt hat Die Bl cke d rfen ERST DANN aus dem Instrument entnommen werden nachdem das Ger t von der Stromversorgung getrennt worden ist und sich...

Page 38: ...Notes...

Page 39: ...we cannot guarantee that interference will not occur in practise Where there is a possibility that injury damage or loss might occur if equipment malfunctions due to radio frequency interference or f...

Page 40: ...les antylia com W coleparmer com Canada T 514 355 6100 E info antylia ca W coleparmer ca Other T 1 847 549 7600 Ordering Information Order No Series Model Legacy SKU 36610 01 BH 200 BH 200 2 SBH130 36...