CALIBRATION
Due to the design of this product, there are no calibrations and/or adjustments to be
performed other than set-ups performed by the customer. Calibration of the Analog I/O
requires a special calibration fixture at the factory.
Additional technical specifications are available in the unit's OPERATORS MANUAL
.
FUNCTIONAL TESTS
1.00 Test equipment required:
1.01 Contact Tachometer (C-P Cat. # 8203-50 or 8203-80)
1.02 Variac 115 VAC, 60Hz or 230 VAC, 50Hz appropriate for unit being
tested.
1.03 Foot switch (C-P Cat. # 7595-42) or Remote Control (C-P Cat. # 7592-83)
2.00 TESTING:
2.01 Plug controller into the variac set to 115 volts ac or 230 volts appropriate
for the unit under test.
2.02 Turn the controllers power switch to the
ON
position. Verify that all LED’s,
display segments, and decimal points illuminate.
2.03 Verifying each LED illuminates for
INT
,
mA
, and
V
, press the
MODE
switch to get to the
INT
mode.
2.04 Select any tubing size (not RPM), then press the
FLOW
key. The