5.2.3  Cleaning the accessories 

Wipe the accessories with a soft cloth or sponge using a neutral detergent, and then with a dry cloth. 

 

 

  Since there is a risk of electric shock, before cleaning, please turn off the power switch and 

disconnect the plug from the outlet.   

  If  the  device  is  contaminated  with  toxic  chemicals  or  gases,  please  make  sure  to  wear 

chemical-resistant gloves and masks when cleaning the interior part. 
 

  Never use chlorine detergent. It may cause the chamber to be oxidized. Also, never use volatile 

substances such as cleaning agents, abrasives, benzene, acids or solvents. 

  Do  not  wipe  the  surface  of  the  instrument  using  an  organic  solvent such  as  sulfuric  acid  or 

hydrochloric acid. 

  Use a soft cloth or a sponge to wash the cleaning tool with a neutral detergent. 

  Do not pour water directly into the instrument while cleaning. 

  Do not arbitrarily disassemble and clean the instrument. 

  If  cleaning  the  instrument  in  a  way  other  than  the  above  method,  please  check  with  Cole-

Parmer or the branch before cleaning. 

 

 

5.3  Major maintenance checks 

Please check the confirmation below once again and operate the device correctly. 

(1) Please check the connection between the device and the power source. 

(2) Please check whether the device and system are correctly fixed in the caster. 

(3) Please check whether the up/down, left/right horizontality of the device is correct. 

(4) Please check if there are any flammable or explosive substances left in the device. 

 

 

 

 

 

 

 

 

 

 

• Do not arbitrarily disassemble or modify the device other than what is described 

in this user manual. 

  When handling electrical components inside the device, it must be handled 

only by qualified people. 

Summary of Contents for 52412-15

Page 1: ...User Manual of Forced Convection Oven Model OVF 800 series 52412Mechanicaloven_ver_1 2...

Page 2: ...t Disclaimer The material in this manual is for the users information only The products it describes are subject to change without prior notice due to the manufacturer s continuous development program...

Page 3: ...reading 11 1 4 1 Warnings 12 1 4 2 Caution 14 2 0 Functional Description 16 2 1 Introduction 17 2 2 Feature 17 2 2 1 Excellent Performance 17 2 2 2 Safety Enhancement 18 2 2 3 Improved Convenience 18...

Page 4: ...uto Tuning 51 4 4 4 Alarm 53 4 4 5 Password 54 4 4 6 Factory Default 54 4 5 Safety Function 55 4 5 1 Safety Functions Related to Temperature 55 4 5 2 Safety Functions Related to Use 55 4 5 3 Alarm Dis...

Page 5: ...of Accessories 68 8 0 Appendix 69 8 1 Technical Specification 70 8 1 1 OVF 800 Series 70 8 2 Disposal of Product 71 8 3 Warranty 71 8 3 1 Warranty period 71 8 3 2 Warranty Exception 71 8 3 3 How to Re...

Page 6: ...52412Mechnicaloven_ver_1 2 6 1 0 Safety...

Page 7: ...atures and structure of the instrument The chapter 3 contains installation instructions for instrument confirmation and installation Also it describes preparation procedures before operation and insta...

Page 8: ...ciency and productivity is to establish effective safety programs and comply with safety rules In addition to maintaining and repairing the instrument regularly please check the condition of the instr...

Page 9: ...the instrument If you need to move the instrument or use it after a long pause make sure that the instrument is operating normally Please stop using the instrument when it appears with the following...

Page 10: ...nts Indicates a hazardous situation which if not avoided will result in death fatal injury or property damage Indicates a hazardous situation which if not avoided could result in serious injury or pro...

Page 11: ...fety Symbols The graphics below are used to convey important information about the use and characteristics of the instrument without further explanation Non safety Symbols Permissible ambient temperat...

Page 12: ...ower connections can cause damage to the instrument and serious personal injury or death Do not handle or touch the electrical cords or electrical components with wet hands 3 Installation Avoid contac...

Page 13: ...nge sounds smells or smoke Do NOT operate damaged or short circuit equipment Do NOT move the instrument when the power cord is connected or when the instrument is operating 5 Repair and Modification D...

Page 14: ...chine Operation When initiating the instrument the operator should check the machine continuously until it reaches to normal status Wear gloves when handling samples to prevent burns since there may b...

Page 15: ...its surroundings Do not disassemble when cleaning the instrument If you are cleaning the instrument in a way other than the above please check the cleaning manner with Cole Parmer 6 Repair The electri...

Page 16: ...52412Mechnicaloven_ver_1 2 16 2 0 Functional Description...

Page 17: ...rs the heat to the samples quickly evenly and precisely 3 A reliability and reproducibility of temperature An excellent reliability and reproducibility of temperature distribution has been verified by...

Page 18: ...een and the alarm sound is activated 6 Instantaneous Outage Recovery Features If the instrument has a short period of power outage occurs within 30 seconds the system returns to the previous operating...

Page 19: ...the vent hole cover 3 Controller Various controllers and electrical components are embedded in the instrument Since the instrument is in excessive heat please wear gloves when controlling the vent ho...

Page 20: ...aker A circuit breaker is embedded to protect the device from short circuits or overcurrent C Power cord socket The socket connects the power line to the instrument The instrument is in excessive heat...

Page 21: ...ttached and the product is manufactured F Over Temperature Limiter If the heat rises excessive temperature the temperature control stops and a warning sign will be displayed on the screen along with a...

Page 22: ...he heater If the door is opened for more than 2 minutes the displays will show you an error message with an alarm sound Moreover the instrument operation will be stopped with the power disconnection w...

Page 23: ...52412Mechnicaloven_ver_1 2 23 3 0 Installation...

Page 24: ...to the instrument and make sure it matches the model you ordered If there is any component below that does not match with your order please contact the branch or device seller of Cole Parmer Please r...

Page 25: ...nstall the instrument in a place with a possibility of flammable gas leakage Do not use or store flammable dust combustible gas around this instrument Operation of the device is prohibited in a place...

Page 26: ...rument on a flat and solid floor for any case of an accident Since the door is opened to the left with maximum 115 please consider sufficient installation space Please keep the appropriate space typic...

Page 27: ...wn in the figure below It is also recommended to install it in an environment equipped with ventilation facilities such as a fume hood or arm hood unit mm Please avoid samples or shelves contacted to...

Page 28: ...he height of the shelf by attaching the shelf support to the shelf adjustor on the side of the chamber as shown below Place the shelf on the shelf support attached to the shelf adjustor Depending on t...

Page 29: ...55 710 C Interior depth 210 330 350 376 D Shelf width 266 346 451 516 E Shelf depth 195 315 335 361 F The height between the top hole of the shelf installation and the inner top 80 90 87 5 85 G Height...

Page 30: ...nformation about potential risks or how to avoid them For safe operation all users must check the safety label before using it and read the 1 4 Instructions for Safe Use required reading carefully to...

Page 31: ...put terminal on the side of the body 3 Connect the other side of the power plug to a power outlet 4 Turn on the power switch and circuit breaker to supply power to the device Before installing the pro...

Page 32: ...n This device is not an explosion proof structure Do not place explosive flammable flammable or oxidative chemicals such as alcohol benzene gasoline on top of the device store them or put them in the...

Page 33: ...there may be residual heat even after the power is off Before using the instrument please make sure the inside of the chamber is clean all times If this instrument has moisture please stop operation i...

Page 34: ...52412Mechnicaloven_ver_1 2 34 4 0 Operation...

Page 35: ...ervation end reservation setting temperature control and stop 4 2 2 General Settings You can check and change the settings of the device such as sound settings date time modification and language sele...

Page 36: ...tton to make general and system settings 2 You can set the target temperature value 3 Warnings When a warning occurs an alarm sign appears and a pop up notification window appears 4 Current Temperatur...

Page 37: ...ot operate with a door open sound Please close the door and restart it If the door is opened while the device is operating the fan and heater will stop operating and a door open warning message will b...

Page 38: ...on without pressing Enter the temperature settings will be canceled The set temperature can be input from 0 to 250 and the temperature control range of the device is Amb 10 250 To set and start the te...

Page 39: ...l of the device after a set time And after reaching the set temperature there is a wait off timer function to control the temperature maintenance for a set time 1 Move to Timer Settings by pressing th...

Page 40: ...imer Set with the unit of hh mm In case setting the start timer as 1 hours 4 You can return to the Basic Mode screen by pressing button with the Wait ON OFF timer applied When the timer setting is com...

Page 41: ...Timer is ON Wait OFF Timer is ON If you only set the Wait ON Timer as ON the instrument waits for the input value before raising the temperature as the set value In this case the heater will be contin...

Page 42: ...etting temperature If you set up the Wait OFF Timer the Wait OFF Timer works automatically once the set temperature has reached In case Wait OFF Timer is set 1 If you press the start button after sett...

Page 43: ...on again If you want to change the Wait OFF timer during timer operation you must turn off the Wait OFF Timer and reset the process 4 2 5 Instantaneous Outage Recovery Function If the power is restore...

Page 44: ...Sound Display Clock Language Auto Run Safety Lock Door Event Communication Pressing the General Settings button in the home screen takes you to the General Settings 4 3 1 Sound Display In Sound Displ...

Page 45: ...n 1 OVF 800 Series does not support the on lamp by door 4 3 2 Clock The Clock is a function that sets the time within the device The time set in the Clock is the standard for the date and time of the...

Page 46: ...orted later 4 3 4 Event Event is the function to indicate the start of control and errors occurred in the device The latest event occurrence and time are displayed Up to 4 events can be displayed on o...

Page 47: ...on It is the function that allows communication settings with LC GreenBox Option Item Function Value Range Initial Value Equipment Address Set communication Address 1 255 1 Baud Rate Set Communication...

Page 48: ...he user sets the Auto Run and the power is restored after the device is turned off for 30 seconds or more during temperature control follow the procedure below Previous States Operation in Basic Contr...

Page 49: ...Temperature Input Calibration Auto Tuning Alarm Password and Factory Default 1 If you press the System Settings button on home screen the password window appears 2 After entering the set password pre...

Page 50: ...e decimal place 0 1 1 1 4 4 2 Calibration When measuring the temperature inside the chamber by using the user s temperature sensor if there is a temperature difference between the temperature sensor i...

Page 51: ...etup Calibration and enter Temperature 1 Temperature 2 and Temperature 3 in the Uncalibrated column Also enter the internal temperature measured by the external temperature sensor in the Calibrated co...

Page 52: ...in program control 3 It moves directly to the operation display window of Basic Mode and the icon on the top status bar blinks indicating that Auto Tuning is running 4 When Auto Tuning is completed t...

Page 53: ...the target temperature allowable upper limit temperature an alarm is generated 0 250 5 Low Deviation Temperature Set the lower limit of temperature deviation If the actual temperature becomes lower t...

Page 54: ...nce If you press the reset button a pop up notification appears and confirms whether or not to proceed with Factory Default This function initializes all settings changed in the Basic Settings General...

Page 55: ...sage for door open appears But there is no alarm sound If the door is closed within the Door Open Alarm Delay time default setting 5 minutes the previous operation continues If the door is not closed...

Page 56: ...is higher than the target temperature allowable upper limit temperature Alarm displays when the actual temperature is lower than the target temperature allowable lower limit temperature Alarm displays...

Page 57: ...uld be stabilized for a certain period of time 3 Turn the red knob counterclockwise using a screwdriver Stop the knob when a warning is displayed on the screen with a notification sound 4 Turn the red...

Page 58: ...52412Mechnicaloven_ver_1 2 58 5 0 Maintenance...

Page 59: ...value Check function of controller Electric wire related checklist Defects if electric wire Check contamination of electric wire Check fixation of wiring connection part Before using the instrument pl...

Page 60: ...t 1 In general Step 1 Take out the shelf from the instrument and wipe it with a dry cloth using a neutral detergent Step2 Wipe the inner surface of the instrument with a dry cloth using a neutral dete...

Page 61: ...h or a sponge to wash the cleaning tool with a neutral detergent Do not pour water directly into the instrument while cleaning Do not arbitrarily disassemble and clean the instrument If cleaning the i...

Page 62: ...rtation 6 When the instrument is not in use for a long period of time unplug the power plug clean it pack it and store it in a dry place Do not move the instrument with the power cord connected Do not...

Page 63: ...52412Mechnicaloven_ver_1 2 63 6 0 Trouble Shooting...

Page 64: ...amaged outlet plug power cord Power supply does not match the equipment s power type Check the voltage phase and capacity of the power supply and connect the power correctly Either circuit breaker is...

Page 65: ...vice failure Request service In case the temperature control is not working Please check whether there is any device that generates strong high frequency noise around the instrument Please remove such...

Page 66: ...the instrument and restart it after the instrument has cooled down If the same problem occurs request service door is open Check if the door text on the display window is on and close the door securel...

Page 67: ...52412Mechnicaloven_ver_1 2 67 7 0 Accessories...

Page 68: ...F 800 26 52412 31 Wire Shelf for OVF 800 61 52412 32 Wire Shelf for OVF 800 112 52412 33 Wire Shelf for OVF 800 147 52412 34 Perforated Shelf for OVF 800 26 52412 35 Perforated Shelf for OVF 800 61 52...

Page 69: ...52412Mechnicaloven_ver_1 2 69 8 0Appendix...

Page 70: ...eel wire electro polished Insulation Mineral wool Ventilation slide Stainless steel dia 38mm x 1ea Dimensions Interior W x D x H mm inch 300x210x400 11 8x8 27x15 75 380x330x480 14 97x13 0x18 9 485x350...

Page 71: ...ratory equipment and its components 8 3 Warranty 8 3 1 Warranty period 1 If the product fails during normal and proper use the warranty period for manufacturing liability shall be two years from the d...

Page 72: ...Details of the Product If you would like to purchase any product accessories please contact the following branch office or head office 8 3 4 Returning Process of the Product If you need to return the...

Page 73: ...E de sales antylia com W coleparmer de T 1 847 549 7600 E sales antylia com W coleparmer com France USA T 33 0 1486 37800 E fr sales antylia com W coleparmer fr T 1 847 549 7600 E sales antylia com W...

Reviews: