52412Mechnicaloven_ver_1.2
41
How to reserve the Wait ON/OFF timer
In case of
Wait ON Timer is OFF / Wait OFF Timer is OFF
In case of
Wait ON Timer is OFF / Wait OFF Timer is ON
- The initial setting is shown as above.
- Timer set does not apply to this instrument.
- If you only set the Wait OFF timer, the
temperature will maintain only for the entered
time once it reaches the target temperature.
- The tempreature control will be terminated if all
the reserved end time is passed.
In case of
Wait ON Timer is ON / Wait OFF Timer is OFF
In case of
Wait ON Timer is ON / Wait OFF Timer is ON
- If you only set the Wait ON Timer as ON, the
instrument waits for the input value before raising
the temperature as the set value.
- In this case, the heater will be continuously
working to maintain the set value until the user
stops the instrument.
- If you turn on both Wait ON Timer and Wait OFF
Timer, you can set the delay time before starting
the heater and also set the maintaining time to
last the set temperature.
- In this case, the heater operation starts after
all the time set in the Wait ON has elapsed. After
reaching the target temperature set by the user,
the target temperature is maintained for the time
set by the Wait OFF. Temperature control is
terminated when all of the set Wait OFF time has
elapsed.
Summary of Contents for 52412-15
Page 1: ...User Manual of Forced Convection Oven Model OVF 800 series 52412Mechanicaloven_ver_1 2...
Page 6: ...52412Mechnicaloven_ver_1 2 6 1 0 Safety...
Page 16: ...52412Mechnicaloven_ver_1 2 16 2 0 Functional Description...
Page 23: ...52412Mechnicaloven_ver_1 2 23 3 0 Installation...
Page 34: ...52412Mechnicaloven_ver_1 2 34 4 0 Operation...
Page 58: ...52412Mechnicaloven_ver_1 2 58 5 0 Maintenance...
Page 63: ...52412Mechnicaloven_ver_1 2 63 6 0 Trouble Shooting...
Page 67: ...52412Mechnicaloven_ver_1 2 67 7 0 Accessories...