52411Vacuumoven0.9_1.9_ver_1.0 as version

 

www.coleparmer.com

                                                 

                                                                                                                     

NOTE: 

the acoustic signal will not end until it is stopped by the operator, but the heating cycle is 

terminated so for the samples inside the instrument will remain exposed to the internal temperature 

the chamber. 

 
 
 

5.4 Delay of Heating Cycle Start 

 

It’s possible to set a delay (hour and minutes) of heating cycle start. 

 

1.

 

Simultaneously Press

 and  

for  few seconds until the time position shows 

LK 0000

 

 

confirm the“0000” password pressing shortly one time

.  

On the top right part of display the 

parameter “dy” (delay) appears close to value 00:00.

 

 

2. Set the desired delay value (hh:mm) pressing

or

 keys. 

It’s possible a quick movement 

between the digits using the

 button. Confirm the set value with another press of 

 button.  

The display comes back to the standby screen. 

3. Pressing the 

 button with long pressure (4-5 seconds) the instrument starts the work cycle but it 

doesn’t immediately heat: the word “end” and the set delay time alternately blink on the top right part 

of display, counting the wait time until the real starting of heating.  
Once the delayed time is passed the instrument starts to heat and the regular timer appears on 
display. 
 

 

5.5. Alarm / Faults 

The oven has a built-

in overtemperature protection circuit. If the oven’s 

temperature overshoots, the 

controller will enter an Alarm Mode that disables all heating elements to protect your samples. 
When in Alarm Mode, a red alarm indicator will display on the screen. The oven will automatically 
resume operation when the temperature has dropped back below the overshoot threshold. 
Setting the temperature setpoint higher will also resume heating function. 

If your oven alarm is sounding due to an elapsed timer, press the Any key to mute! 
 
 
 
 
 
 
 
 

Summary of Contents for 52411-22

Page 1: ...52411Vacuumoven0 9_1 9_ver_1 0 as version www coleparmer com Vacuum Oven Model 52411 22 23 24 25 User Manual...

Page 2: ...User Warning WARNING Failure to follow all warnings and instructions could result in serious injury Your safety is very important to us so we urge you to take the following precautions when using thi...

Page 3: ...e Model Wattage 52411 22 550W 52411 24 650W 52411 23 550W 52411 25 650W Cord Connected Models The unit will be supplied with a mains lead fitted with either US EU UK or Indian plug Should the lead not...

Page 4: ...Oven Front Fig 2 Oven Rear 2 Components Vacuum Gauge inHg Vacuum Release Temperature Controller Light Power Switch Vacuum Valve Inert Gas purge Valve Door Lock Handle Shelves Standard Circuit Breaker...

Page 5: ...F 1O C 1 8O F 1O C Electrical Characteristics 120 VAC 220 VAC 60 Hz 50 Hz 550W 550W 120 VAC 220 VAC 60 Hz 50 Hz 650 W 650W Gasket Silicone Silicone Fuse Circuit breaker 120V 10A Circuit Breaker 220V 5...

Page 6: ...instructions please see the next section g To release vacuum upon completion first power off the oven a Rotate the Vacuum Release Valve to release your vacuum using atmospheric pressure Or b Connect...

Page 7: ...quickly change the digit Program units tens etc of the value of the parameter you are editing 3 In combination with the key allows access to menus with password 4 and adjustment buttons allow users to...

Page 8: ...ed on pressingone time the button the set temperature value starts to blink Set the desired temperature value in Celsius degrees pressing keys 2 It s possible a quick movement between the digits using...

Page 9: ...The display comes back to the standby screen 3 Pressing the button with long pressure 4 5 seconds the instrument starts the work cycle but it doesn t immediately heat the word end and the set delay ti...

Page 10: ...faulty Oven significantly overheats Oven constantly heating Temperature set too high See section 5 1 Temperature Setting Faulty heating relay Replace if R 1M Inaccurate temperature reading Controller...

Page 11: ...m gauge needle does not move door can open easily The door is not closed well The door does not create a complete seal between gasket and glass Loosen nut 3 5 turns counterclockwise Tighten door pin 1...

Page 12: ...measurement than the controller Please note laser probe measurements will be inaccurate without compensating for emissivity Refer to your laser probe s user manual 1 Simultaneously Press and for few s...

Page 13: ...Vernon Hills IL 60061 T 800 323 4340 E sales coleparmer com W coleparmer com EU Antylia Scientific GmbH Futtererstr 16 97877 Wertheim Deutschland T 49 9377 9203 0 E cde sales antylia com W coleparmer...

Reviews: