52411Vacuumoven0.9_1.9_ver_1.0 as version
6.1.
Calibration
While the oven arrives pre-calibrated from the factory, certain adjustments may be necessary for optimal
performance. Please note these adjustments will affect your oven settings! We recommend writing down
any factory settings prior to adjustment so they may be reset in the event of undesirable operation.
Probe Calibration
Follow the instructions below to adjust the temperature readout. These steps are only necessary in the
event there is a discrepancy between the display and measurements taken with an alternate device.
For example: a thermometer placed inside the oven with a different measurement than the controller.
Please note: laser probe measurements will be inaccurate without compensating for emissivity. Refer to
your laser probe’s user manual.
1. Simultaneously Press
and
for few seconds until the time position shows
“
LK 0000
”
2.
Press
to set LK 0003
3.
Press
three times, you can see Pb, this is Temperature offset on single point.
4.
After set the Pb value
,
Press
to exit.
Check elevation chart to
confirm maximum possible
vacuum
A pipe connection or
fitting is loose
Tighten and secure vacuum
connections
Check release dial and purge
valve
The vacuum gauge is
damaged
Replace vacuum gauge
Oven vacuum leak
Vacuum loss of >1”Hg
within 24 hours
Door gasket is worn
Replace door gasket
Valves improperly closed
Check valve positions
A pipe connection or
fitting is loose
Tighten and secure vacuum
connections
Vacuum valve is leaking
Replace vacuum valve