![COKEVA 11940C Operation Note Download Page 13](http://html.mh-extra.com/html/cokeva/11940c/11940c_operation-note_2635063013.webp)
Cokeva
11940C
Operation Note
8
Instrument
Critical
Specifications
Recommended
Model
Spectrum
Analyzer
Frequency
Range:
30
MHz
–
1
GHz
HP
8567A
(10
kHz
–
1.5
GHz)
HP
8568B
(100
Hz
–
1.5
GHz)
HP
8566B
(100
Hz
–
22
GHz)
Tracking
Generator
Frequency
Range:
30
MHz
–
1
GHz
HP
8444A,
Option
059
(100
MHz
–
1.5
GHz)
For
use
with
HP
8567A
or
HP
8568B
Signal
Generator
Frequency
Range:
30
MHz
–
1
GHz
HP
8656B
(0.1
–
990
MHz)
Preamplifier*
Frequency
Range:
30
MHz
–
1
GHz
RF
Gain:
20
dB
HP
8447D
(100
kHz
–
1.3
GHz)
(Gain
>26
dB)
*
HP
85685A
RF
Preselector
may
be
substituted
if
a
spectrum
analyzer
listed
above
is
used.
Table
3.
Recommended
Test
Setup
Equipment
3
‐
17.
Measurement
Techniques
3
‐
18.
The
probe
is
designed
to
be
held
very
close
to
potential
radiating
points.
The
plane
of
the
probe
tip
loops
must
be
perpendicular
to
the
radiating
magnetic
field
as
illustrated
in
Figure
2.
For
example,
if
the
seam
of
an
enclosure
is
being
tested
for
EMl,
hold
the
probe
as
illustrated
in
Figure
7.
When
measuring
the
radiation
from
a
coaxial
cable,
position
the
probe
tip
as
shown
in
Figure
8.
The
side
of
the
probe
with
the
calibration
label
should
be
facing
away
from
the
cable
being
measured.
3
‐
19.
The
probe
may
be
used
as
a
source
in
susceptibility
testing
as
described
in
Section
3
‐
11
and
Paragraph
3
‐
16.
In
this
application,
hold
the
probe
within
1
cm
of
the
electronic
circuit
component
being
tested.
4
‐
1.
OPERATION
VERIFICATION
4
‐
2.
Operation
Verification
for
the
Cokeva
11940C
consists
of
the
Return
Loss
Measurement
Test
included
in
this
section.
Actual
antenna
factor
verification
requires
a
calibrated
broadband
magnetic
field.
A
return
loss
measurement
provides
a
reliable
alternate
method
of
determining
the
probe's
electrical
integrity.
The
test
is
performed
over
a
frequency
range
where
the
VSWR
is
approximately
1.25:1
for
a
properly
operating
probe.
Electrical
damage
is
indicated
by
any
large
decrease
in
return
loss
(increase
in
VSWR)
as
specified
in
the
test.
4
‐
3.
Probes
that
fail
this
test,
although
damaged,
may
still
detect
electromagnetic
radiation.
In
this
case,
the
antenna
factors
listed
on
the
unit
are
no
longer
valid
and
the
circuitry
no
longer
rejects
electric
field
coupling.
Recalibration
of
the
probe's
antenna
factors
is
not
possible.