Coherent Solutions SwitchPXIe 1000 Series User Manual Download Page 27

PXIe Platform Optical Switch Module

|

 

SwitchPXIe

 

Coherent Solutions Ltd. 

Version 

2.03

 

26 

 

Common System Command Descriptions 

 

Command 

*CLS

 

Syntax 

*CLS 

Description 

Clear Status command  

Parameters 

None 

Response 

None 

Example 

*CLS 

 

Command 

*IDN?

 

Syntax 

*IDN? 

Description 

Query the chassis identification  

Parameters 

None 

Response 

Comma separated string with the <manufacturer>,<server name>,<chassis controller 
name>,<server version> 

Example 

*IDN? -> Coherent Solutions,CSLServer,PXIE-8133,FW2.0.15 

 

Command 

*OPC?

 

Syntax 

*OPC? 

Description 

Query the Operation Complete Status  

Parameters 

None 

Response 

1

 

is returned if all the modules installed in the chassis are ready to execute 

 

commands 

0

 

is returned if any module installed in the chassis still has a command to 

 

execute in the input queue

 

Example 

*OPC? -> 0 

 

Command 

*OPT?

 

Syntax 

*OPT? 

Description 

Query the modules managed by the CSLServer  

Parameters 

None 

Response 

Response will be a comma separated string of the installed modules in the chassis 

Example 

*OPT? -> ,SwitchPXIe-1004-1-FA,LaserPXIe-1001-4-FC,,VOAPXIe-1001-2-FA
 

,,,,O2EPXIe-1001-1-FC,,,,,,,,, 

 

 

Summary of Contents for SwitchPXIe 1000 Series

Page 1: ...SwitchPXIe 1000 1100 1400 Series Optical Switch USER MANUAL...

Page 2: ...y result from its use No license is granted by implication or otherwise under any patent rights of Coherent Solutions The information contained in this publication is subject to change without notice...

Page 3: ...1xN models 10 4 Connecting Optical Input 11 Cleaning and Connecting Optical Fibers 11 5 Handling the SwitchPXIe Module 12 SwitchPXIe Module Installation 12 SwitchPXIe Module Uninstallation 14 6 CSL So...

Page 4: ...on 2 03 iii 9 6 1 NI MAX 30 9 6 2 SCPI Multi Chassis Commands 31 10 Maintenance 32 11 Technical Support 35 Contacting the Technical Support Group 35 Transportation 35 12 Warranty 36 General Informatio...

Page 5: ...Do not proceed unless the required conditions are met and understood CAUTION Indicates a potentially hazardous situation which if not avoided may result in minor or moderate injury Do not proceed unl...

Page 6: ...res other than those specified herein may result in exposure to hazardous situations or impair the protection provided by this unit CAUTION The SwitchPXIe modules are sensitive to electrostatic discha...

Page 7: ...e and low loss switching to various test set ups while taking advantage of the PXIe mainframe s flexible connectivity and high channel density form factor This manual applies to SwitchPXIe 1000 1100 1...

Page 8: ...cator LEDs 3 Red Optical ports 4 Blue Optical ports 5 PXIe Module Optical Connector Information 6 Fastening clip 7 SwitchPXIe Module Information 8 PXIe Headers IMPORTANT All Crossover switch modules D...

Page 9: ...Output Optical port 1 4b Blue Input Output Optical port 1 5a Red Input Output Optical port 2 5b Blue Input Output Optical port 2 6 PXIe Module Optical Connector Information 7 Fastening clip 8 PXIe He...

Page 10: ...Overview Note Number of common and routing ports will depend on the selected model SwitchPXIe 1106 shown above 1 Fastening screw 2 Error Status LED 3 Common Optical port 4 Routing Optical ports 5 PXI...

Page 11: ...2 For SwitchPXIe Crossover models The SwitchPXIe XX04 is a 2x2 crossover switch This means that it has only 2 valid states Transmit T and Cross C Solid green Indicates which state the SwitchPXIe is i...

Page 12: ...pyl alcohol b Use compressed air to dry completely c Visually inspect the fiber end to ensure its cleanliness 3 Carefully align the connector and port to prevent the fiber end from touching the outsid...

Page 13: ...ease be sure to wear a grounded wrist strap at all times when handling the SwitchPXIe module to prevent such damage Take care not to handle the connectors on the SwitchPXIe module as once they are exp...

Page 14: ...module are properly connected to the chassis 6 Secure the module into the chassis by pulling up on the black locking clip A click should be heard when the locking clip is properly engaged into the fr...

Page 15: ...enough time to finish boot procedures and initialize the communication server 1 Shutdown the PXIe chassis 2 Unfasten the locking clip on the SwitchPXIe module Unscrew the fastening screw Pull out the...

Page 16: ...oller to communicate with the SwitchPXIe installed in the chassis some software packages are necessary These software packages are all contained in the single installer CSLDriverInstaller Drivers for...

Page 17: ...ltiple Chassis Mode additional hardware modules are required The Chassis Mode can be changed at any time so it is recommended to select Single Mode until all other configuration requirements have been...

Page 18: ...ons be ticked to allow cohesionUI to work on any network configuration IMPORTANT 5 Continue with the installation of the CSLDriverInstaller by following the on screen installation prompts 6 Once compl...

Page 19: ...to the PXIe chassis via an ethernet connection To obtain the IP address open the Command Prompt window on the chassis controller and then run the ipconfig command Note down the IPv4 address that is di...

Page 20: ...select a module from the list that is displayed Controls for a specific module or a channel in a module are then displayed Settings You can use the SETTINGS page to configure the units or step size u...

Page 21: ...frequency or power increments decrements when the or button is clicked To set the step size click appropriate values on the settings page Info Each time when the Info button is clicked a cohesionUI in...

Page 22: ...etween HIDDEN or SHOWN The default mode is to hide the empty slots Update cohesionUI with the latest information from the CSL server HOME Click the synchronization button next to the title of the chas...

Page 23: ...Modules menu button on the left will bring up a dropdown that the SwitchPXIe module can also be selected from After clicking the desired SwitchPXIe module its control page is displayed All informatio...

Page 24: ...d port click the Port number eg PORT 2 The SwitchPXIe can route the light very quickly however cohesionUI takes a few seconds to update the displayed routing Therefore there is a slight delay between...

Page 25: ...e CSLServer Parameter Default Unit Alternative Units Power DBM DBM Frequency HZ THZ GHZ MHZ KHZ Frequency Fine HZ THZ GHZ MHZ KHZ Wavelength NM M PM Argument Data Format wsp Specifies whitespace chara...

Page 26: ...module this is an integer 1 to 18 m is the channel index of a specific unit in the module this is an integer 1 2 Message Queues Information is exchanged in the form of messages These messages are held...

Page 27: ...oherent Solutions CSLServer PXIE 8133 FW2 0 15 Command OPC Syntax OPC Description Query the Operation Complete Status Parameters None Response 1 is returned if all the modules installed in the chassis...

Page 28: ...omplete bit Query the modules installed on the slot Query the Identifier for the slot returns the manufacturer part number serial number hardware and firmware versions Configuration commands Descripti...

Page 29: ...ameters None Response The response will be a comma separated string of detectors installed in the SwitchPXIe If a module is not installed in a channel it will not return any identification string Exam...

Page 30: ...the error status of the specified channel of the SwitchPXIe Parameters None Response 1 Switch reporting an error 0 No error on switch Example ROUT1 CHANA ERROR 0 Command ROUTe n CHANnel m STATE Syntax...

Page 31: ...PXIe 8834 to PXIe 8381 connection Chassis 2 has the controller running CSLServer and Chassis 3 is the extended chassis IMPORTANT The CSLServer does not manage the chassis numbers These are controlled...

Page 32: ...is will always return 1 Example In Single chassis mode SYSTEM CHASSIS 1 SYSTEM CHASSIS LIST 0 SYSTEM CHASSIS MODE SINGLE In Multi chassis mode SYSTEM CHASSIS 2 SYSTEM CHASSIS LIST 2 3 SYSTEM CHASSIS M...

Page 33: ...rature in a clean and dry area Keep the unit out of direct sunlight Avoid high humidity or significant temperature fluctuations Avoid unnecessary shocks and vibrations If any liquids are spilled on or...

Page 34: ...cates that this toxic or hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirement in SJ T11363 2006 SJ T11363 2006 Part Name Toxic o...

Page 35: ...PXIe Platform Optical Switch Module SwitchPXIe Coherent Solutions Ltd Version 2 03 34 KCC INFORMATION Product KCC Registration No Logo SwitchPXIe XXXX XX R R Csn SwitchPXIe...

Page 36: ...nd the serial number see the product identification label as well as a description of your problem close at hand You may also be requested to provide software and module version numbers This informati...

Page 37: ...IAL DAMAGES Liability Coherent Solutions shall not be liable for damages resulting from the use of the product nor shall be responsible for any failure in the performance of other items to which the p...

Page 38: ...n Merchandise Authorization RMA form that we will send to you We will also provide an RMA number and shipping instructions 3 Pack the equipment in its original shipping material Be sure to supply a st...

Page 39: ...eneral enquiries info coherent solutions com Technical support support coherent solutions com Tel 64 9 478 4849 www coherent solutions com www linkedin com company coherent solutions ltd www facebook...

Reviews: