Status LED
The Status LEDs are used to denote the operation state of the SwitchPXIe, and will indicate
which state the switch is in.
The different colors have slightly different meanings for the different SwitchPXIe models.
3.5.1
For SwitchPXIe 1x1 and 1x4 models:
•
Blinking yellow -
Indicates initialization on startup.
•
All off -
Indicates that the SwitchPXIe is in the
PARK
state (
State 0
).
•
Solid green -
Indicates which state the SwitchPXIe is routed to.
•
Solid red -
Indicates a Switch error. Most often this means an invalid route position.
3.5.2
For SwitchPXIe Crossover models:
The SwitchPXIe-XX04 is a 2x2 crossover switch. This means that it has only 2 valid states:
Transmit (T) and Cross.(C)
•
Solid green -
Indicates which state the SwitchPXIe is in (default is T)
•
Solid red -
Indicates there was an error in switching states. Only 1 LED will light up.
3.5.3
For SwitchPXIe Duplex models:
The SwitchPXIe-XX05 is a duplex switch. This means that it has only 2 valid states: 1 and 2.
•
Solid green
–
Indicates which state the SwitchPXIe is in (default is 1)
•
Blinking red -
Indicates there was an error in switching states. Only 1 LED will light up.
3.5.4
For SwitchPXIe 1xN models:
•
Blinking yellow -
Indicates initialization on startup.
•
Solid green -
Indicates successful initialization and communication.
•
Solid red -
Indicates a Switch error. Most likely an initialization error at startup.