2. Enter the
Rmax
and
Rmin
values from the calibration card.
3. Select the
Center Target
button. A live image appears in the Calibration screen. Center the verifier over the
Master Symbol on the Data Matrix symbol of your traceable calibration card.
4. Once the symbol is centered in the field of view, select
Start Calibration
.
15
Setting Up Your DataMan 475 Verifier