WARNINGS
The BioXp™ 3200 system is heavy. Do not attempt to lift
or move the instrument without the assistance and/or
supervision of colleagues. Be certain to use appropriate
moving equipment and proper lifting techniques (including but
not limited to making certain that you have a secure grip on
the instrument before lifting, keeping your spine in a neutral
position while lifting with your legs, and coordinating lifting and
moving movements with all appropriate personnel). Improper
lifting can cause permanent back injury.
Uncrating instructions
1. Remove the quarter turn latches from the crate and lift off
the sides and top.
Figure 1
2. Remove the plastic bag from the top of the unit and pull
down to expose the lower portion of the unit and remove
the accessory box and upper foam support.
3. If possible, lift the pallet approximately 24” off the ground
with a forklift and set it on a sturdy base.
4. Have a movable cart ready which is capable of holding
greater than 200 lbs.
5. Using two people, lift the unit clear of the base and set on
the movable cart.
6. Move the system using the cart to the prepared site.
7. Codex recommends keeping the shipping materials until
installation and acceptance testing have been completed.
8. Remove tape holding lid and open lid.
9. Remove any tape internal to the system holding the
thermal cycler door closed.
10. Remove the two white covers for the chillers from the
accessory box and place them in the storage locations
(shown in figure 2).
Figure 2: Thermal cover placement
11. Once in the final location, the shipping screw needs to
be removed which is located under the pipette head.
Caution:
Use care when removing this screw as sharp
edges exist under the head (shown in figure 3).
Figure 3: Shipping screw location (shown with head cover removed)
12. After removing the power cord from packaging, connect it
to the power inlet on the system cart located on the right
rear side panel.
13. Check that the power switch located above the power
cord inlet is in the “off” position (0).