Codex BioXp 3200 Uncrating And Crating Download Page 2

Page 2 of 3
Part number 45027  

 Revision 1.1  

|

  Effective 17APR2020

© 2020 Codex DNA, Inc. All rights reserved.

 WARNINGS

The BioXp™ 3200 system is heavy. Do not attempt to lift 
or move the instrument without the assistance and/or 
supervision of colleagues. Be certain to use appropriate 
moving equipment and proper lifting techniques (including but 
not limited to making certain that you have a secure grip on 
the instrument before lifting, keeping your spine in a neutral 
position while lifting with your legs, and coordinating lifting and 
moving movements with all appropriate personnel). Improper 
lifting can cause permanent back injury.

Uncrating instructions

1.  Remove the quarter turn latches from the crate and lift off 

the sides and top.

  

Figure 1

2.  Remove the plastic bag from the top of the unit and pull 

down to expose the lower portion of the unit and remove 
the accessory box and upper foam support.

3.  If possible, lift the pallet approximately 24” off the ground 

with a forklift and set it on a sturdy base. 

4.  Have a movable cart ready which is capable of holding 

greater than 200 lbs.  

5.  Using two people, lift the unit clear of the base and set on 

the movable cart.

6.  Move the system using the cart to the prepared site.

7.  Codex recommends keeping the shipping materials until 

installation and acceptance testing have been completed. 

8.  Remove tape holding lid and open lid.

9.  Remove any tape internal to the system holding the 

thermal cycler door closed.

10.  Remove the two white covers for the chillers from the 

accessory box and place them in the storage locations 
(shown in figure 2).

 

Figure 2: Thermal cover placement

11.  Once in the final location, the shipping screw needs to 

be removed which is located under the pipette head. 

Caution: 

Use care when removing this screw as sharp 

edges exist under the head (shown in figure 3). 

 

  

Figure 3: Shipping screw location (shown with head cover removed)

12.  After removing the power cord from packaging, connect it 

to the power inlet on the system cart located on the right 
rear side panel. 

13.  Check that the power switch located above the power 

cord inlet is in the “off” position (0). 

Reviews: