19
NUCLEUS® 7 HYBRID MODE PROFESSIONALS GUIDE
TR
OUBLESHOO
T
Observations, possible causes and solutions
PROBLEM
POSSIBLE CAUSE
TEST / RESOLUTION
Recipient can’t
hear anything
through the
Hybrid system
Batteries are flat
Check batteries and replace/recharge if necessary.
Wax is blocking a
filter
1. Replace the dome (with its built-in wax filter).
2. Check the wax filter on the speaker unit and
replace if necessary.
Educate the recipient to do this regularly.
Hybrid receiver is
broken
Replace the Hybrid receiver.
Sound processor
fault
1. Perform a listening test using a Stetoclip. If this
shows a problem, send the sound processor for
repair.
2. Perform a listening test using a Cochlear Monitor
Earphone Adaptor and earphones. If this shows a
problem, send the sound processor for repair.
Feedback
The Hybrid receiver
cable is not the right
length
1. Check the Hybrid receiver cable length. If it is too
long, the cable will not sit nicely against the ear
and the speaker unit may sit too low in the ear
canal. If it is too short, the dome may keep
falling out. Adjust so it sits snugly and without
tension. See
Shape the Hybrid receiver cable
on
page 22.
2. If it is still not correct, change to a different
length Hybrid receiver.
Dome vent is too
large
Change to a dome with less venting.
Dome fit is too loose
Change to a bigger dome.
The sound
processor LED
lights steady
orange
Incorrect Hybrid
receiver is connected
Change to the correct Hybrid receiver.