18
NUCLEUS® 7 HYBRID MODE PROFESSIONALS GUIDE
Troubleshoot
This chapter provides troubleshooting procedures for the Hybrid system, only for clinician-
oriented tasks.
For product-specific troubleshooting, see:
• the
Nucleus 7 Sound Processor User Guide
• the
Nucleus 7 Hybrid User Guide
• the
Custom Sound
®
Software User Guide
or the online help.
Equipment required
Check regularly that the items listed below are in good repair and well stocked. Have
plenty of spare parts – especially receivers, domes, earhooks and wax filters.
• Nucleus 7 Hybrid Fitting Kit
• Cochlear EAC200 Series Stetoclip
• Cleaning set, wax loop, tissues
• Replacement
ProWax
filters
• Vent
cleaner
• Cochlear EAC200 Series Screwdriver.
Troubleshooting procedure
1. Interview
the
recipient.
2. Take an otoscopic examination and ensure there is no wax blockage in the ear canal.
3. Inspect the sound processor thoroughly.
4. Ensure that thresholds have not changed by repeating pure tone audiometry. Identify
any middle ear issues with Otoscopy and Tympanometry.
5. Refer to the following table for suggested solutions.