24
COCHLEAR CHARGERS USER GUIDE
Other information
Electromagnetic compatiblity (EMC)
Guidance and manufacturer’s declaration – electromagnetic
emissions
The Cochlear Home Charger or Cochlear Portable Charger are
intended for use in the electromagnetic environment specified
below. The customer or the user of the chargers should assure
they are used in such an environment.
EMISSIONS
TEST
COMPLIANCE ELECTROMAGNETIC
ENVIRONMENT – GUIDANCE
RF emissions
CISPR 11
Group 1
The Cochlear Home Charger and
Cochlear Portable Charger use RF
energy only for internal functions.
Therefore, RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
Class B
(Cochlear Home Charger Charging
Mode, Cochlear Portable Charger
Body-Worn Charging Mode)
The Cochlear Home Charger and
Cochlear Portable Charger are suitable
for use in all establishments, including
domestic and those directly connected
to the public low voltage power supply
network that supplies buildings used
for domestic purposes.
The Cochlear Home Charger and
Cochlear Portable Charger are suitable
for use in clinics and hospitals.
Harmonic
emissions
IEC 61000-3-2
Not applicable* Not applicable*
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable* Not applicable*
* Chargers do not connect to mains supply. Power adapter supplied is a
commercially available part.