HANDBOOK (V
4.0)
COMPLETE CONTROL SYSTEM GEN2
52
There are no lights or sounds coming from the Coapt Complete Calibrate button.
1.
Ensure all wired connections to the prosthetic device(s) are correct according to the Coapt Device Cable
Connection Guide for the cable type of your prosthesis (e.g. “H0”, “I4”, etc.).
2.
Ensure the prosthesis’ battery is charged and the prosthesis is turned on.
3.
If the Coapt system will connect via Bluetooth to the
Complete ControlRoom
application, the
Complete
Calibrate
button unit may be damaged and subject for replacement.
4.
If the Coapt system will NOT connect via Bluetooth to the
Complete ControlRoom
application, the Coapt
system does not appear to be receiving power. Ensure the power connection with the prosthesis is correct.
FUNCTION/CALIBRATION
Control of the Virtual Arm in the Coapt application is fine but the prosthesis doesn’t move.
1.
Make sure “Device” is selected on the toggle in the Actuate panel of the
Complete ControlRoom
application.
2.
Ensure all wired connections to the prosthetic device(s) are correct according to the Coapt Device Cable
Connection Guide for the cable type of your prosthesis (e.g. “H0”, “I4”, etc.)
3.
Ensure all of the settings on the other manufacturers’ devices are set exactly as shown in the Coapt
guidance documentation specific to the components connected by the cable type of your prosthesis
(e.g. “H0”, “I4”, etc.)
The [specific function/motion] is not working correctly.
1.
Ensure the prosthesis’ battery is charged and the prosthesis is turned on.
2.
Ensure all wired connections to the prosthetic device(s) are correct according to the Coapt Device Cable
Connection Guide for the cable type of your prosthesis (e.g. “H0”, “I4”, etc.)
3.
Ensure all of the settings on the other manufacturers’ devices are set exactly as shown in the Coapt
guidance documentation specific to the components connected by the cable type of your prosthesis
(e.g. “H0”, “I4”, etc.)
The prosthesis is moving on its own when I’m wearing it and relaxed.
1.
Verify in the
Complete ControlRoom
application that none of your EMG inputs are showing warnings
for electrode lift-off. Address any that are.
2.
It is possible that the Coapt system sees your data for that motion (the one that is moving when you are
relaxed) as very similar to your data for the actual relax state. A few things that may help:
a.
Calibrate the motion that is moving when you don’t want it to with muscle contractions that are a
little harder/stronger.
b.
Calibrate “No Motion” while being sure not to be making muscle contractions.