– 75 –
input port ± 2,0 kV
IEC 61000-4-5 (2006): Surge immunity test
Test levels on AC Power input port: AC Power input port, ±0,5
kV and ±1,0 kV differential mode.
IEC 61000-4-6 (2014): Immunity to conducted disturbances in
the frequency range 0,15 – 80 MHz
Test levels on AC Power input port: 6 Vrms with 80 % AM @ 2
Hz.
IEC 61000-4-8 (2009): Immunity to power frequency magnetic
fi
elds
Test level: 30 A/m, 50 Hz and 60 Hz
IEC 61000-4-11 (2004): Voltage Dips and Interruptions
Test levels on AC Power input port:
Test frequency: 50 Hz. Test voltages: 100 and 230 V.
100 % for
½
cycle positive and negative half period.
60 % for 10 cycles.
30 % for 25 cycles.
100 % for 250 s.