15
– Service and Maintenance
T34L™
Operator Manual
26
15
– Service and Maintenance
T34L
TM
Syringe Pump requires annual maintenance and calibration. In between maintenance the Pump
requires only battery charging and cleaning between patients (or as necessary). It is recommended that
the performance of the Pump is checked periodically.
Warning:
If the Pump is damaged in anyway return to an authorized service centre.
Cleaning
Before connecting the pump to a patient, and periodically during use, clean the unit using a lint-free cloth
lightly dampened with warm water and a mild detergent, disinfectant or 10% bleach solution.
Once a month (or as required) clean the main Pump screw thread and guiding rods with a small
dry brush to remove debris or other particles.
Warning:
Always turn the syringe Pump off before cleaning.
Caution:
Do not clean the
T34L
TM
syringe pump with chemicals such as Xylene, Acetone or
similar solvents. These chemicals can cause damage to components and labels. Use a lint-free
cloth dampened with warm water and mild detergent or 10% bleach solution.
Caution:
Do not soak or immerse any part of the
T34L
TM
in water or any other solution.
Storage
If the pump is to be stored for an extended period it should be cleaned. Store in a clean, dry atmosphere
at room temperature and, if available, use the original packaging, or a suitable alternative, for protection.
.
Service
The Pump must only be serviced or repaired by one of the following:
CME or its approved service representative,
Qualified biomedical technicians trained and certified.
Caution:
Refer all service, repair and adjustments only to qualified & certified technical
personnel. Unauthorized modifications or the use of any spare parts, other than those supplied
by the manufacturer or their distributor, will void any warranty.
NOTE:
DISPOSAL/DECOMISSIONING -
when the time comes to dispose of the pump,
accessories or packaging do so in the best way to minimize any negative impact on the
environment. You may be able to use special recycling or disposal schemes. To find out about
these contact your local waste disposal service. Existing national or local regulations concerning
waste disposal must take precedence over the above advice.