Technical   Specifications   

   

   

Sensor   kind   

Analog,   generates   an   output   voltage   between   

0 - 5 V 

   

Measuring   range      

0   ..   14      

Resolution   using      
12-­‐   bits   5V   AD   converter   

   
0.005   pH   

Temperature   range   

5°C   and   80°C   

Isopotential   pH   

pH   7   (point   at   which   temperature   has   no   effect   on   output)   

Calibration   function   

pH   =   -­‐4.04   *   V

out

   (V)+   13.68   

Connection   

IEEE1394   connector   for   BT-­‐IEEE1394   sensor   cable.      
Sensor   cable   not   delivered   with   the   sensor.      

   

   

   

   

   

   

   

   

Warranty:      

The    pH    Sensor    BT61i    is    warranted    to    be    free    from    defects    in    materials    and   
workmanship   for   a   period   of   12   months   from   the   date   of   purchase   provided   that   it   has   
been    used    under    normal    laboratory    conditions.    This    warranty    does    not    apply    if    the   
sensor   has   been   damaged   by   accident   or   misuse.      

   

Note:

   

This   product   is   to   be   used   for   educational   purposes   only.   It   is   not   appropriate   for   

industrial,   medical,   research,   or   commercial   applications.              

   

   

 

Rev. 20