background image

2   |   

BT61i   pH   Sensor   User’s   Guide      

Short   description   

The    CMA    pH    Sensor    is    a    general-­‐purpose    pH    measurement    system    that    allows   
measuring   the   degree   of   acidity/pH   value   of   a   solution   in   the   range   between   0   and   14.   
The   sensor   consists   of   a   pH   amplifier   and   pH   electrode.   The   pH   Electrode   (031)   is   not   
delivered   with   the   pH   sensor   and   has   to   be   purchased   separately.      

   

The   pH   Sensor   can   be   directly   connected   to   analog   BT   inputs   of   the   CMA   interfaces.   
The   sensor   cable   BT   -­‐   IEEE1394   needed   to   connect   the   sensor   to   an   interface   is   not   
supplied   with   the   sensor   and   has   to   be   purchased   separately   (CMA   Article   BTsc_1).      

Sensor   recognition      

The   pH   Sensor   has   a   memory   chip   (EEPROM)   with   information   about   the   sensor:   its   
name,    measured    quantity,    unit    and    calibration.    Through    a    simple    protocol    this   
information   is   read   by   the   CMA   interfaces   and   the   sensor   is   automatically   recognized   
when   it   is   connected   to   these   interfaces.         
If   your   pH   Sensor   is   not   automatically   detected   by   an   interface   you   have   to   manually   
set   up   your   sensor   by   selecting   it   from   the   Coach   Sensor   Library.      

pH   Amplifier      

The    CMA    pH    amplifier    is    a    device,    which    allows    a   
standard    combination    pH    electrode    (such    as    the   
CMA   pH   Electrode   (031))   to   be   monitored   by   a   lab   
interface.   The   pH   electrode   is   connected   to   the   pH   
amplifier   via   BNC   connector   located   on   one   end   of   
the   sensor   box.   The   pH   amplifier   adjusts   the   voltage   
produced   by   the   pH   electrode   to   a   range   between   0   
to   5   V,   which   can   be   measured   by   an   interface.   
   

pH   Electrode         

The    CMA    pH    electrode    is    designed    to    make   
measurements    in    the    pH    range    of    0    to    14.      
It    has    a    coax    cable,    with    a    BNC    connector.      
The   electrode   (made   from   glass)   is   built   in   a   plastic   
tube    with    an    opening    at    the    bottom    side.    It    is   
supplied   in   a   bottle   filled   with   a   protective   solution.   
When   the   pH   electrode   is   not   being   used,   it   must   be   
kept   in   this   liquid.   For   use   and   maintenance   of   the   
CMA   pH   electrode   follow   instructions   given   in   its   
User’s   Guide.      

Figure   1.

   The   pH   Amplifier   (BT61i).      

Figure   2.   

The   pH   Electrode   (031).   